Beyond Mean Bias in Automated ECG Interval Measurement: A Record-Level Audit in PTB-XL+

This archive provides the submission manuscript, algorithm specification, deidentified derived analysis tables, figures, analysis manifest, and core Python scripts for a record-level audit of automated ECG interval measurements in PTB-XL+. The custom Tp-e delineator returned accepted outputs for 102 of 160 ECGs and failed all four algorithm-performance acceptance criteria. Among accepted outputs, substantial disagreement with the reader-derived human reference remained. The archive separates explicit non-output from error among retained outputs and includes calibration-overlap sensitivity analyses. Interpretation boundary: the evaluated delineator must not be used for confirmatory clinical-effect analyses. The package does not establish clinical risk thresholds or treatment effects. No credentials, private sample-mapping key, local reader-source CSVs, or unnecessary raw-waveform copies are included. PTB-XL and PTB-XL+ remain subject to their upstream terms.

Authors

Institutions

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2026-11-17
DOI
https://doi.org/10.5281/zenodo.24311922
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Beyond Mean Bias in Automated ECG Interval Measurement: A Record-Level Audit in PTB-XL+

Yilong Hao, P. Andy Li, Hui Ye, Song Chen et al.
1 links
Zenodo (CERN European Organization for Nuclear Research)
ECG Monitoring and Analysis
article

Beyond Mean Bias in Automated ECG Interval Measurement: A Record-Level Audit in PTB-XL+

Yilong Hao, P. Andy Li, Hui Ye, Song Chen, Jian Wan, Qian Zhang
article en

Abstract

This archive provides the submission manuscript, algorithm specification, deidentified derived analysis tables, figures, analysis manifest, and core Python scripts for a record-level audit of automated ECG interval measurements in PTB-XL+. The custom Tp-e delineator returned accepted outputs for 102 of 160 ECGs and failed all four algorithm-performance acceptance criteria. Among accepted outputs, substantial disagreement with the reader-derived human reference remained. The archive separates explicit non-output from error among retained outputs and includes calibration-overlap sensitivity analyses. Interpretation boundary: the evaluated delineator must not be used for confirmatory clinical-effect analyses. The package does not establish clinical risk thresholds or treatment effects. No credentials, private sample-mapping key, local reader-source CSVs, or unnecessary raw-waveform copies are included. PTB-XL and PTB-XL+ remain subject to their upstream terms.

Zenodo (CERN European Organization for Nuclear Research)
Shanghai Jiao Tong University (CN), Shanghai University of Traditional Chinese Medicine (CN), Seventh People's Hospital of Shanghai (CN), Pudong New Area People's Hospital (CN)
Good health and well-being
ECG Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon