Modeling of novel converging beam radiotherapy system in Geant4 Monte Carlo simulation

PURPOSE: A Geant4-based model was created to simulate radiation transport within a novel radiotherapy system developed by Convergent Radiotherapy and Radiosurgery (Convergent RnR; Haifa, Israel). This treatment device incorporates a Bragg-diffraction X-ray lens to focus X-rays from a conventional X-ray tube towards a small but finite focal spot with quasi-monoenergetic photons. METHODS: The simulation model included an X-ray tube and a CRnR focusing lens consisting of single-crystal aluminum tiles. To validate the Geant4 model of this device, transverse and longitudinal in-air and in-water dose distributions from the converging beam field were calculated and compared to radiochromic film measurements. RESULTS: The implemented Monte Carlo (MC) simulation model predicted a mean photon energy of 53.96 keV and showed agreement within 0.5% when compared with the SpekPy calculation. Both the simulation and experiment produced an attenuation characteristics of 3.2 mm Al. In-water tissue maximum ratios showed percentage differences ≤3.03% at depths of 5, 10, 30, 40, 50, 60, and 70 mm. Exceptions were observed at 20, 80, and 85 mm, likely due to film positioning uncertainty. A gamma analysis of in-water scanning dose distributions demonstrated a >96% pass rate for the 2%/2 mm criterion and >99% for the 3%/3 mm criterion across all focal depths. CONCLUSIONS: In this study, CRnR's converging beam radiotherapy system was implemented in Geant4 to demonstrate the device operation and characterize its physical and dosimetric outputs. The MC simulation implemented in this work successfully demonstrated converging dose distributions with Bragg-diffraction physics, which were in good agreement with radiochromic film measurements. This simulation model may be used to further refine the converging beam system design for clinical implementation, including validation of the treatment planning system.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-11-17
DOI
https://doi.org/10.1088/2057-1976/aec2f8
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Modeling of novel converging beam radiotherapy system in Geant4 Monte Carlo simulation

Leonard Che Fru, Benjamin Abraham Insley, Aviad Schori, Reza Reiazi et al.
1 links
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

Modeling of novel converging beam radiotherapy system in Geant4 Monte Carlo simulation

Leonard Che Fru, Benjamin Abraham Insley, Aviad Schori, Reza Reiazi, Michael Kleckner, Dirk Alan Bartkoski, Dongyeon Lee, Mohammad Salehpour, Ramesh C. Tailor
article en

Abstract

PURPOSE: A Geant4-based model was created to simulate radiation transport within a novel radiotherapy system developed by Convergent Radiotherapy and Radiosurgery (Convergent RnR; Haifa, Israel). This treatment device incorporates a Bragg-diffraction X-ray lens to focus X-rays from a conventional X-ray tube towards a small but finite focal spot with quasi-monoenergetic photons. METHODS: The simulation model included an X-ray tube and a CRnR focusing lens consisting of single-crystal aluminum tiles. To validate the Geant4 model of this device, transverse and longitudinal in-air and in-water dose distributions from the converging beam field were calculated and compared to radiochromic film measurements. RESULTS: The implemented Monte Carlo (MC) simulation model predicted a mean photon energy of 53.96 keV and showed agreement within 0.5% when compared with the SpekPy calculation. Both the simulation and experiment produced an attenuation characteristics of 3.2 mm Al. In-water tissue maximum ratios showed percentage differences ≤3.03% at depths of 5, 10, 30, 40, 50, 60, and 70 mm. Exceptions were observed at 20, 80, and 85 mm, likely due to film positioning uncertainty. A gamma analysis of in-water scanning dose distributions demonstrated a >96% pass rate for the 2%/2 mm criterion and >99% for the 3%/3 mm criterion across all focal depths. CONCLUSIONS: In this study, CRnR's converging beam radiotherapy system was implemented in Geant4 to demonstrate the device operation and characterize its physical and dosimetric outputs. The MC simulation implemented in this work successfully demonstrated converging dose distributions with Bragg-diffraction physics, which were in good agreement with radiochromic film measurements. This simulation model may be used to further refine the converging beam system design for clinical implementation, including validation of the treatment planning system.

Biomedical Physics & Engineering Express
The University of Texas MD Anderson Cancer Center (US), The University of Texas System (US), Convergent Science (United States) (US), Convergent Engineering (United States) (US), MD Anderson Cancer Center Madrid (ES), Anderson University - South Carolina (US)
Good health and well-being
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon