A Compact Eight-Parameter Analytical Model of Photon-Beam Lateral Dose Profiles: Plateau, Penumbra and Out-of-Field Tail Decomposition for Elekta Linear Accelerators

To develop and validate a compact analytical model of megavoltage photon-beam lateral dose profiles, with eight physically interpretable parameters per profile, for beam-data quality assurance (QA) and treatment planning system (TPS) commissioning support. The relative dose at off-axis position x, for a field of half-width R, is expressed as D(x,R) = A·F(x,R)·P(x,R) + S(x,R), where F (a fourth-order polynomial in x/R) describes the plateau, P (an error-function pair of width σ) the penumbra, and S (an exponential) the out-of-field tail. The parameters were fitted to 58 in-plane profiles measured on an Elekta linac at 6, 10 and 15 MV (1×1 to 40×40 cm², depths 1.5-20 cm), and the model was tested on 14 independent profiles (second linac, Monte Carlo). Agreement was assessed with the root-mean-square error (RMSE), R² and the global gamma index at 2%/2 mm and 1%/1 mm. All profiles passed 2%/2 mm at 100%, with a mean RMSE of 0.551% (range 0.24-1.19%) and R² ≥ 0.9978; the independent profiles gave a mean RMSE of 0.541%. At 1%/1 mm, 99.2% of points passed, and all 6 MV profiles and all fields ≥ 4×4 cm² passed at ≥ 95%; the remaining deviations were confined to small 10-15 MV fields and low-gradient regions. The parameters followed physically consistent trends: σ increased with depth and, for fields ≥ 5×5 cm², with energy, B increased with depth and with field size up to 30×30 cm², α decreased with depth, and H and H₄ tracked the loss of horns with depth. Each parameter retains a direct physical meaning, and the parameters recovered simulated jaw-calibration and steering errors set at or below the AAPM TG-142 tolerances, supporting their use in beam-data QA and commissioning; independent dose calculation will require a continuous parameterization in depth and field size.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-11-17
DOI
https://doi.org/10.1088/2057-1976/aec2f0
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A Compact Eight-Parameter Analytical Model of Photon-Beam Lateral Dose Profiles: Plateau, Penumbra and Out-of-Field Tail Decomposition for Elekta Linear Accelerators

Dikra Bakari, Mohammed Rezzoug, Ibrahim Ait Ouaggou, Abdeslem Rrhioua et al.
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
article

A Compact Eight-Parameter Analytical Model of Photon-Beam Lateral Dose Profiles: Plateau, Penumbra and Out-of-Field Tail Decomposition for Elekta Linear Accelerators

Dikra Bakari, Mohammed Rezzoug, Ibrahim Ait Ouaggou, Abdeslem Rrhioua, Mustapha Zerfaoui, Yassine Oulhouq, Samir Didi, Hassan Ou Hadda, Omar Hamzaoui
article en

Abstract

To develop and validate a compact analytical model of megavoltage photon-beam lateral dose profiles, with eight physically interpretable parameters per profile, for beam-data quality assurance (QA) and treatment planning system (TPS) commissioning support. The relative dose at off-axis position x, for a field of half-width R, is expressed as D(x,R) = A·F(x,R)·P(x,R) + S(x,R), where F (a fourth-order polynomial in x/R) describes the plateau, P (an error-function pair of width σ) the penumbra, and S (an exponential) the out-of-field tail. The parameters were fitted to 58 in-plane profiles measured on an Elekta linac at 6, 10 and 15 MV (1×1 to 40×40 cm², depths 1.5-20 cm), and the model was tested on 14 independent profiles (second linac, Monte Carlo). Agreement was assessed with the root-mean-square error (RMSE), R² and the global gamma index at 2%/2 mm and 1%/1 mm. All profiles passed 2%/2 mm at 100%, with a mean RMSE of 0.551% (range 0.24-1.19%) and R² ≥ 0.9978; the independent profiles gave a mean RMSE of 0.541%. At 1%/1 mm, 99.2% of points passed, and all 6 MV profiles and all fields ≥ 4×4 cm² passed at ≥ 95%; the remaining deviations were confined to small 10-15 MV fields and low-gradient regions. The parameters followed physically consistent trends: σ increased with depth and, for fields ≥ 5×5 cm², with energy, B increased with depth and with field size up to 30×30 cm², α decreased with depth, and H and H₄ tracked the loss of horns with depth. Each parameter retains a direct physical meaning, and the parameters recovered simulated jaw-calibration and steering errors set at or below the AAPM TG-142 tolerances, supporting their use in beam-data QA and commissioning; independent dose calculation will require a continuous parameterization in depth and field size.

Biomedical Physics & Engineering Express
Mohamed I University (MA), Centre Hospitalo-Universitaire Mohammed VI d'Oujda (MA)
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon