Computational Fluid Dynamics Analysis of Percutaneous left ventricular assist device Support Modes on Ventricular Hemodynamics and Hemocompatibility

Percutaneous left ventricular assist devices (PLVADs) provide cardiac support in heart failure, yet the hemodynamic effects of varying support levels and the balance between ventricular unloading and blood trauma remain unclear. This study employed numerical simulations to analyze a PLVAD-supported left ventricle under three pump speeds: low, high, and normal control. Analysis encompassed both conventional hemodynamic parameters (e.g., TAWSS, OSI, RRT) and validated mathematical model. Results showed that PLVAD support enlarged the vortex core and increased vorticity in the left ventricle, while reducing them at the aortic outlet. It effectively lowered ventricular wall pressure-achieving unloading-but also decreased aortic outlet WSS. Furthermore, PLVAD elevated OSI and ECAP while reducing TAWSS and RRT, effects that intensified at higher speeds. Increased speed also raised non-physiological scalar shear stress (SSS) and platelet activation, significantly increasing the hemolysis index in the rotational domain and elevating blood damage risk in the ventricular chamber compared to normal physiology. In conclusion, while PLVAD effectively unloads the ventricle, it impairs flow pulsatility and hemocompatibility. Controlling high-speed-induced SSS and platelet activation is critical to mitigate hemolysis. Monitoring hemocompatibility indicators-especially in the apical and aortic outlet regions-is essential due to their link with thrombosis and adverse events.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2026-11-17
DOI
https://doi.org/10.1088/1873-4030/aec2ec
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Computational Fluid Dynamics Analysis of Percutaneous left ventricular assist device Support Modes on Ventricular Hemodynamics and Hemocompatibility

Yuan He, Qilian Xie, Yuehu Wu, Honglong Yu et al.
Medical Engineering & Physics
Mechanical Circulatory Support Devices
article

Computational Fluid Dynamics Analysis of Percutaneous left ventricular assist device Support Modes on Ventricular Hemodynamics and Hemocompatibility

Yuan He, Qilian Xie, Yuehu Wu, Honglong Yu, Fengqing Zhang, 陈增生, Haibo Chen, Xuefeng Feng, Xianqiang Wang
article en

Abstract

Percutaneous left ventricular assist devices (PLVADs) provide cardiac support in heart failure, yet the hemodynamic effects of varying support levels and the balance between ventricular unloading and blood trauma remain unclear. This study employed numerical simulations to analyze a PLVAD-supported left ventricle under three pump speeds: low, high, and normal control. Analysis encompassed both conventional hemodynamic parameters (e.g., TAWSS, OSI, RRT) and validated mathematical model. Results showed that PLVAD support enlarged the vortex core and increased vorticity in the left ventricle, while reducing them at the aortic outlet. It effectively lowered ventricular wall pressure-achieving unloading-but also decreased aortic outlet WSS. Furthermore, PLVAD elevated OSI and ECAP while reducing TAWSS and RRT, effects that intensified at higher speeds. Increased speed also raised non-physiological scalar shear stress (SSS) and platelet activation, significantly increasing the hemolysis index in the rotational domain and elevating blood damage risk in the ventricular chamber compared to normal physiology. In conclusion, while PLVAD effectively unloads the ventricle, it impairs flow pulsatility and hemocompatibility. Controlling high-speed-induced SSS and platelet activation is critical to mitigate hemolysis. Monitoring hemocompatibility indicators-especially in the apical and aortic outlet regions-is essential due to their link with thrombosis and adverse events.

Medical Engineering & Physics
Fu Wai Hospital (CN), Beihang University (CN)
National Natural Science Foundation of China, National Key Research and Development Program of China
Good health and well-being
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon