Calibration of 3‐T MR Relaxometry for Quantitative Liver Iron Assessment Using Human Liver Explants

MR relaxometry has become the preferred noninvasive method for estimating hepatic iron from R2 and R2* relaxation rates. However, the precise quantification of liver iron concentration (LIC) requires calibration, and published equations relating relaxation rates to LIC have limited transferability across scanners and acquisition protocols. This study aimed to establish local 3-T calibration equations relating R2 and R2* relaxation rates to chemically determined LIC using human liver explants and to evaluate their applicability in independent in vivo cohorts. Thirty-six human liver explants underwent MR examination on a 3-T system using multi-echo spin-echo and multi-echo gradient-echo sequences to obtain R2 and R2* relaxation rates. LIC was determined chemically by atomic absorption spectrophotometry. Calibration equations were derived using linear regression and compared with previously published models. The R2*-based calibration was subsequently applied to independent in vivo cohorts examined using the same MR system and comparable acquisition protocols. Both MR methods correlated strongly with chemically determined LIC (R2*: r = 0.95, p < 0.001; R2: r = 0.92, p < 0.001). Compared with published models, the local calibrations demonstrated lower prediction errors and narrower limits of agreement within the studied explant cohort. Application of the R2* equation to in vivo datasets yielded physiologically plausible LIC estimates, with the lowest values observed in patients with primary sclerosing cholangitis, intermediate values in healthy controls and the highest values in subjects with elevated serum ferritin concentrations. Explant-derived 3-T calibration equations demonstrated strong association with chemically measured LIC and generated physiologically plausible LIC estimates in vivo. These findings support the feasibility of deriving local MR relaxometry calibrations from human liver explants, while further prospective validation against an independent LIC reference is required before clinical application.

Authors

Institutions

Publication Details

Journal
NMR in Biomedicine
Published
2026-11-16
DOI
https://doi.org/10.1002/nbm.70440
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Calibration of 3‐T MR Relaxometry for Quantitative Liver Iron Assessment Using Human Liver Explants

Iva Subhanová, Dita Pajuelo, Petr Šedivý, Martin Květoň et al.
NMR in Biomedicine
Advanced MRI Techniques and Applications
article

Calibration of 3‐T MR Relaxometry for Quantitative Liver Iron Assessment Using Human Liver Explants

Iva Subhanová, Dita Pajuelo, Petr Šedivý, Martin Květoň, Ondřej Fabián, Milan Hájek, Monika Dezortová, Ferdinand Los, Peter Kordač
article en

Abstract

MR relaxometry has become the preferred noninvasive method for estimating hepatic iron from R2 and R2* relaxation rates. However, the precise quantification of liver iron concentration (LIC) requires calibration, and published equations relating relaxation rates to LIC have limited transferability across scanners and acquisition protocols. This study aimed to establish local 3-T calibration equations relating R2 and R2* relaxation rates to chemically determined LIC using human liver explants and to evaluate their applicability in independent in vivo cohorts. Thirty-six human liver explants underwent MR examination on a 3-T system using multi-echo spin-echo and multi-echo gradient-echo sequences to obtain R2 and R2* relaxation rates. LIC was determined chemically by atomic absorption spectrophotometry. Calibration equations were derived using linear regression and compared with previously published models. The R2*-based calibration was subsequently applied to independent in vivo cohorts examined using the same MR system and comparable acquisition protocols. Both MR methods correlated strongly with chemically determined LIC (R2*: r = 0.95, p < 0.001; R2: r = 0.92, p < 0.001). Compared with published models, the local calibrations demonstrated lower prediction errors and narrower limits of agreement within the studied explant cohort. Application of the R2* equation to in vivo datasets yielded physiologically plausible LIC estimates, with the lowest values observed in patients with primary sclerosing cholangitis, intermediate values in healthy controls and the highest values in subjects with elevated serum ferritin concentrations. Explant-derived 3-T calibration equations demonstrated strong association with chemically measured LIC and generated physiologically plausible LIC estimates in vivo. These findings support the feasibility of deriving local MR relaxometry calibrations from human liver explants, while further prospective validation against an independent LIC reference is required before clinical application.

NMR in BiomedicineVol. 39(12)
Charles University (CZ), Institute of Clinical and Experimental Medicine (CZ), Thomayer University Hospital (CZ), General University Hospital in Prague (CZ)
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon