Curvature-Based Morphotypes of the Distal Radius for Volar Plate Selection: A Computational Analysis of 100 Radii
Purpose To characterize the 3-dimensional morphology of the distal radius and to test whether plate-to-bone conformity differs between volar locking plate designs and depends on individual morphology. Methods In this retrospective, computational, computed tomography−based morphometric study, 100 radiologically unaffected distal radii from 100 adult patients were analyzed. Longitudinal curvature, transverse central volar prominence, and distal width were quantified on a standardized anatomical coordinate system. Five plate designs (Synthes, Intercus, Medartis, Arthrex, and Bonebridge) were simulated on each radius, yielding 500 plate-to-bone pairings. Plate pose was optimized computationally to maximize apposition without bone penetration. Contact area was defined as the proportion of plate surface within 0.1 mm of the bone. A curvature-based selection rule was derived by maximizing the mean contact area and evaluated by using leave-one-out cross-validation and permutation testing. Results The mean contact area ranged from 15.9% to 18.2% across the five designs, with no significant difference between them ( P = .80), and no design was disproportionately represented among the individually best-fitting plates ( P = .65). Almost all variation in contact area was attributable to the radius rather than to the plate design. Morphology differed between sexes, whereas conformity did not ( P = .95). The plate design interacted with longitudinal curvature ( P = .028): conformity increased on flatter radii for Bonebridge and decreased for Arthrex. The transverse central volar prominence and distal width showed no interaction ( P = .13 and .96). A selection rule based on longitudinal curvature improved the mean contact area from 18.2% to 20.2% under cross-validation. Conclusions The distal radius morphology varies substantially between individuals, and this variation exceeds the differences between commercially available volar locking plate designs. Longitudinal curvature showed a small design-specific effect on conformity, whereas transverse concavity and distal width did not. The proposed selection rule is exploratory and requires prospective validation before clinical use. Type of study/level of evidence Therapeutic IV.
Authors
- Bastian Sigrist (ORCID: https://orcid.org/0000-0002-7720-1420)
- Lisa Reissner (ORCID: https://orcid.org/0000-0003-4291-3759)
- Andreas Schweizer (ORCID: https://orcid.org/0000-0003-0609-7467)
- Yichao Luo (ORCID: https://orcid.org/0009-0005-4556-2458)
Institutions
- University of Zurich (CH)
- Universitätsklinik Balgrist (CH)
Publication Details
- Journal
- Journal of Hand Surgery Global Online
- Published
- 2026-11-17
- DOI
- https://doi.org/10.1016/j.jhsg.2026.101199
- Primary Topic
- Orthopedic Surgery and Rehabilitation
- Type
- article
- Field-Weighted Citation Impact
- 0.00