Analytical evaluation of a urine immunoassay for 2‐ethylidene‐1,5‐dimethyl‐3,3‐diphenyl pyrrolidine ( EDDP ) with focus on cutoff optimization

Presented is an analytical evaluation of the quantILab Methadone Metabolite (EDDP) immunoassay (Werfen, Italy) performed on the ILab Taurus clinical chemistry analyzer, with assessment of three cutoff concentrations (100, 200, and 300 ng/mL) using liquid chromatography-tandem mass spectrometry (LC-MS/MS) as the confirmatory reference method. A total of 100 human urine samples (50 EDDP-positive and 50 EDDP-negative by routine screening) were analyzed and compared with a confirmatory LC-MS/MS method. Confirmatory LC-MS/MS analysis showed EDDP concentrations ranging from 48.8 to 25,500 ng/mL. At a cutoff of 100 ng/mL, the immunoassay achieved a sensitivity of 100% and a specificity of 98%, with no false-negative results observed. Increasing the cutoff to 200 and 300 ng/mL resulted in a slight decrease in sensitivity (98%) and the occurrence of false-negative results, while specificity reached 100%. Overall accuracy remained 99% across all tested cutoff values. The occurrence of discordant results was limited to specimens with EDDP concentrations close to the evaluated cutoff values, whereas samples with higher EDDP concentrations were consistently classified correctly. These findings demonstrate that cutoff selection significantly influences the diagnostic performance of immunoassay-based screening. In forensic toxicology, where false negatives are particularly critical, a cutoff of 100 ng/mL appears to provide the most appropriate balance between sensitivity and specificity. The evaluated immunoassay represents a reliable tool for high-throughput screening, although confirmatory analysis by LC-MS/MS remains essential.

Authors

Institutions

Publication Details

Journal
Journal of Forensic Sciences
Published
2026-11-16
DOI
https://doi.org/10.1111/1556-4029.70517
Primary Topic
Forensic Toxicology and Drug Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Analytical evaluation of a urine immunoassay for 2‐ethylidene‐1,5‐dimethyl‐3,3‐diphenyl pyrrolidine ( EDDP ) with focus on cutoff optimization

Stefania Ortu, Luca Morini, Brian Rossi, Claudia Vignali et al.
Journal of Forensic Sciences
Forensic Toxicology and Drug Analysis
article

Analytical evaluation of a urine immunoassay for 2‐ethylidene‐1,5‐dimethyl‐3,3‐diphenyl pyrrolidine ( EDDP ) with focus on cutoff optimization

Stefania Ortu, Luca Morini, Brian Rossi, Claudia Vignali, Cristiana Stramesi, Valentina Martini
article en

Abstract

Presented is an analytical evaluation of the quantILab Methadone Metabolite (EDDP) immunoassay (Werfen, Italy) performed on the ILab Taurus clinical chemistry analyzer, with assessment of three cutoff concentrations (100, 200, and 300 ng/mL) using liquid chromatography-tandem mass spectrometry (LC-MS/MS) as the confirmatory reference method. A total of 100 human urine samples (50 EDDP-positive and 50 EDDP-negative by routine screening) were analyzed and compared with a confirmatory LC-MS/MS method. Confirmatory LC-MS/MS analysis showed EDDP concentrations ranging from 48.8 to 25,500 ng/mL. At a cutoff of 100 ng/mL, the immunoassay achieved a sensitivity of 100% and a specificity of 98%, with no false-negative results observed. Increasing the cutoff to 200 and 300 ng/mL resulted in a slight decrease in sensitivity (98%) and the occurrence of false-negative results, while specificity reached 100%. Overall accuracy remained 99% across all tested cutoff values. The occurrence of discordant results was limited to specimens with EDDP concentrations close to the evaluated cutoff values, whereas samples with higher EDDP concentrations were consistently classified correctly. These findings demonstrate that cutoff selection significantly influences the diagnostic performance of immunoassay-based screening. In forensic toxicology, where false negatives are particularly critical, a cutoff of 100 ng/mL appears to provide the most appropriate balance between sensitivity and specificity. The evaluated immunoassay represents a reliable tool for high-throughput screening, although confirmatory analysis by LC-MS/MS remains essential.

Journal of Forensic Sciences
University of Pavia (IT)
Forensic Toxicology and Drug Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon