Photon-Counting CT for Postoperative Imaging of Cochlear Implants: Comparison to Flat-Detector CT

OBJECTIVES: After FDA approval in 2021, photon-counting detector computed tomography (PCCT) has demonstrated its feasibility for high-resolution imaging while lowering the radiation dose at the same time. High-resolution visualization of cochlear implant (CI) integrity and positioning is crucial for postoperative management. The aim of this study was to evaluate the performance of PCCT for postoperative imaging after CI implantation in comparison to high-resolution flat-detector CT (FDCT), the current clinical gold standard. MATERIALS AND METHODS: Postoperative PCCT scans after cochlear implantation were retrospectively analyzed from August 2024 to April 2026. Scans were included for analysis if FDCT acquisition had been before PCCT on the same patient in the course of sequential bilateral cochlear implantation. Image quality was assessed in consensus by 2 experienced readers using a scoring system analyzing structural delineation of the inner and middle ear, CI integrity, metal and motion artifacts as well as calculating an overall score. Radiation dose was retrieved from the picture archiving and communication system. RESULTS: Twenty-eight postoperative PCCT scans of patients after CI implantation with a mean time delay of 4.5±5.1 years between FDCT and PCCT scans were evaluated. Image quality of PCCT was noninferior to FDCT for all single imaging features (P<0.001, eg, distinction of single CI electrodes: median [interquartile range] PCCT 3 [3 to 3]; FDCT 3 [3 to 3]; P<0.001) with the exception of the homogeneity of bony structures and the mastoid cell septation (P=0.999). PCCT was superior to FDCT in reducing metal and motion artifacts and increasing the delineation of the stapes (P=0.009 to <0.001; eg, metal artifacts PCCT 3 [2.75 to 3]; FDCT 2 [1 to 2]; P<0.001). Effective radiation dose of PCCT was 0.05±0.01 mSv and 8.8±5.4-times lower than for FDCT with 0.44±0.29 mSv (P<0.001) applying our institutional protocols. CONCLUSIONS: Regarding image quality in postoperative assessment of CI, PCCT is noninferior to or in single features better than FDCT at a significantly lower radiation dose and shorter scan time.

Authors

Institutions

Publication Details

Journal
Investigative Radiology
Published
2026-11-16
DOI
https://doi.org/10.1097/rli.0000000000001320
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Photon-Counting CT for Postoperative Imaging of Cochlear Implants: Comparison to Flat-Detector CT

Jessica Graef, Markus Kopp, Sarina K. Mueller, Lava Taha et al.
5 links
Investigative Radiology
Advanced X-ray and CT Imaging
article

Photon-Counting CT for Postoperative Imaging of Cochlear Implants: Comparison to Flat-Detector CT

Jessica Graef, Markus Kopp, Sarina K. Mueller, Lava Taha, Maximilian Hinsen, Martin Georg Zeilinger, Claudius Sebastian Mathy, Ulrich Hoppe, Joachim Albert Hornung, Michael Uder, Marco Wiesmueller, Stefan Lang, Arnd Dörfler
article en

Abstract

OBJECTIVES: After FDA approval in 2021, photon-counting detector computed tomography (PCCT) has demonstrated its feasibility for high-resolution imaging while lowering the radiation dose at the same time. High-resolution visualization of cochlear implant (CI) integrity and positioning is crucial for postoperative management. The aim of this study was to evaluate the performance of PCCT for postoperative imaging after CI implantation in comparison to high-resolution flat-detector CT (FDCT), the current clinical gold standard. MATERIALS AND METHODS: Postoperative PCCT scans after cochlear implantation were retrospectively analyzed from August 2024 to April 2026. Scans were included for analysis if FDCT acquisition had been before PCCT on the same patient in the course of sequential bilateral cochlear implantation. Image quality was assessed in consensus by 2 experienced readers using a scoring system analyzing structural delineation of the inner and middle ear, CI integrity, metal and motion artifacts as well as calculating an overall score. Radiation dose was retrieved from the picture archiving and communication system. RESULTS: Twenty-eight postoperative PCCT scans of patients after CI implantation with a mean time delay of 4.5±5.1 years between FDCT and PCCT scans were evaluated. Image quality of PCCT was noninferior to FDCT for all single imaging features (P<0.001, eg, distinction of single CI electrodes: median [interquartile range] PCCT 3 [3 to 3]; FDCT 3 [3 to 3]; P<0.001) with the exception of the homogeneity of bony structures and the mastoid cell septation (P=0.999). PCCT was superior to FDCT in reducing metal and motion artifacts and increasing the delineation of the stapes (P=0.009 to <0.001; eg, metal artifacts PCCT 3 [2.75 to 3]; FDCT 2 [1 to 2]; P<0.001). Effective radiation dose of PCCT was 0.05±0.01 mSv and 8.8±5.4-times lower than for FDCT with 0.44±0.29 mSv (P<0.001) applying our institutional protocols. CONCLUSIONS: Regarding image quality in postoperative assessment of CI, PCCT is noninferior to or in single features better than FDCT at a significantly lower radiation dose and shorter scan time.

Investigative Radiology
Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), Universitätsklinikum Erlangen (DE)
Good health and well-being
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon