A Method to Reduce the Factors Affecting Iodine Enhancement for Computed Tomography Contrast Optimization

PURPOSE: Iodine contrast enhancement is assessed using computed tomography (CT) number, though this value is not proportional to iodine concentration. This study develops a metric for iodine enhancement that is directly proportional to iodine concentration. MATERIALS AND METHODS: Using studies previously reported in the literature, we parameterized the effect to create a correction for tube potential and patient-size-specific beam hardening on iodine CT number decrease. To validate the correction, we performed a study using different iodine concentrations, phantom sizes, and tube potentials. We also applied the correction to patient pulmonary angiogram data. Patients were grouped by tube potential [80, 100, 120, 140] (kV). Quantitative iodine contrast enhancement (Q-ICE) (ie, the difference between noncontrast and contrast-enhanced CT number) of the pulmonary artery (PA) was calculated. Pearson correlation coefficients are reported to characterize trends in PA Q-ICE across weight before and after correction. RESULTS: The range of CT numbers measured for [10, 15] mg/mL iodine inserts was [290, 433] HU before correction, and [24, 36] HU after correction. Before correction, median PA Q-ICE enhancement was [662, 490, 263, 175] HU for [80, 100, 120, 140] kV, with a range of 487 HU observed across all measurements. After correction, the median PA Q-ICE enhancement was [366, 373, 285, 249] HU, with a range of 124 HU. Pearson correlation coefficients significantly (P < 0.05) decreased after correction from -0.67 to -0.37. CONCLUSIONS: This study demonstrated a method to normalize effects of tube potential and beam hardening on CT number, yielding a metric proportional to iodine concentration.

Authors

Institutions

Publication Details

Journal
Journal of Thoracic Imaging
Published
2026-11-16
DOI
https://doi.org/10.1097/rti.0000000000000918
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A Method to Reduce the Factors Affecting Iodine Enhancement for Computed Tomography Contrast Optimization

Giuseppe V. Toia, Zahra Alyani Nezhad, Timothy P. Szczykutowicz, Carrie M. Bartels et al.
Journal of Thoracic Imaging
Advanced X-ray and CT Imaging
article

A Method to Reduce the Factors Affecting Iodine Enhancement for Computed Tomography Contrast Optimization

Giuseppe V. Toia, Zahra Alyani Nezhad, Timothy P. Szczykutowicz, Carrie M. Bartels, Rachel M. Bladorn, Sean D. Rose
article en

Abstract

PURPOSE: Iodine contrast enhancement is assessed using computed tomography (CT) number, though this value is not proportional to iodine concentration. This study develops a metric for iodine enhancement that is directly proportional to iodine concentration. MATERIALS AND METHODS: Using studies previously reported in the literature, we parameterized the effect to create a correction for tube potential and patient-size-specific beam hardening on iodine CT number decrease. To validate the correction, we performed a study using different iodine concentrations, phantom sizes, and tube potentials. We also applied the correction to patient pulmonary angiogram data. Patients were grouped by tube potential [80, 100, 120, 140] (kV). Quantitative iodine contrast enhancement (Q-ICE) (ie, the difference between noncontrast and contrast-enhanced CT number) of the pulmonary artery (PA) was calculated. Pearson correlation coefficients are reported to characterize trends in PA Q-ICE across weight before and after correction. RESULTS: The range of CT numbers measured for [10, 15] mg/mL iodine inserts was [290, 433] HU before correction, and [24, 36] HU after correction. Before correction, median PA Q-ICE enhancement was [662, 490, 263, 175] HU for [80, 100, 120, 140] kV, with a range of 487 HU observed across all measurements. After correction, the median PA Q-ICE enhancement was [366, 373, 285, 249] HU, with a range of 124 HU. Pearson correlation coefficients significantly (P < 0.05) decreased after correction from -0.67 to -0.37. CONCLUSIONS: This study demonstrated a method to normalize effects of tube potential and beam hardening on CT number, yielding a metric proportional to iodine concentration.

Journal of Thoracic Imaging
University of Wisconsin–Madison (US), Henry Ford Hospital (US)
Good health and well-being
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon