Evaluation of Contamination Meter Efficiency for Detecting Surface Activity Across Decay Modes and Contaminated Area Sizes in Clinical Settings

BACKGROUND: Use of open sources in nuclear medicine facilities demands reliable surface contamination measurements. Regulatory limits for contaminated areas are set at very low surface activity values, from 0.4 Bq cm-2 to 4 Bq cm-2. OBJECTIVE: This study evaluates the sensitivity of two different contamination meters to detect low contamination values in clinical settings using radionuclides with different decay modes. Additionally, the effect of contaminated area size on sensitivity and quantitative accuracy was assessed. METHODS: Two types of contamination meters were used to measure four radionuclides with different decay modes: 223Ra, 177Lu, 99mTc, and 18F. The measurements were performed at two different distances (2 cm and 10 cm) and over two surface areas (1 cm2 and 16 cm2). RESULTS: At a 2 cm distance and 16 cm2 surface area, the lowest detectable surface activity was 2 Bq cm-2 for alpha- and beta-emitting radionuclides, whereas for 99mTc it was above 1 kBq cm-2. Quantitative accuracy was the highest for 177Lu with surface activity between 5.0-12.5 Bq cm-2, ranging from 1.1% to 20.9%. For the smaller surface area (1 cm2), detection limits were substantially higher: 1 MBq cm-2 for 99mTc and 40 Bq cm-2 for 223Ra, 177Lu, and 18F. CONCLUSION: Measurement sensitivity and accuracy depend strongly on radionuclide decay mode and contaminated area size. Small, concentrated contaminations are detectable only at sufficiently high surface activity whereas lower surface activity values become detectable when the size of the contaminated area increases. Importantly, contamination meters demonstrated optimal performance for alpha and beta emitters, which represent the most significant contamination risk in clinical practice.

Authors

Institutions

Publication Details

Journal
Health Physics
Published
2026-11-16
DOI
https://doi.org/10.1097/hp.0000000000002226
Primary Topic
Radioactive Decay and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Evaluation of Contamination Meter Efficiency for Detecting Surface Activity Across Decay Modes and Contaminated Area Sizes in Clinical Settings

Mikko A. Hakulinen, Heidi Grohn
Health Physics
Radioactive Decay and Measurement Techniques
article

Evaluation of Contamination Meter Efficiency for Detecting Surface Activity Across Decay Modes and Contaminated Area Sizes in Clinical Settings

Mikko A. Hakulinen, Heidi Grohn
article en

Abstract

BACKGROUND: Use of open sources in nuclear medicine facilities demands reliable surface contamination measurements. Regulatory limits for contaminated areas are set at very low surface activity values, from 0.4 Bq cm-2 to 4 Bq cm-2. OBJECTIVE: This study evaluates the sensitivity of two different contamination meters to detect low contamination values in clinical settings using radionuclides with different decay modes. Additionally, the effect of contaminated area size on sensitivity and quantitative accuracy was assessed. METHODS: Two types of contamination meters were used to measure four radionuclides with different decay modes: 223Ra, 177Lu, 99mTc, and 18F. The measurements were performed at two different distances (2 cm and 10 cm) and over two surface areas (1 cm2 and 16 cm2). RESULTS: At a 2 cm distance and 16 cm2 surface area, the lowest detectable surface activity was 2 Bq cm-2 for alpha- and beta-emitting radionuclides, whereas for 99mTc it was above 1 kBq cm-2. Quantitative accuracy was the highest for 177Lu with surface activity between 5.0-12.5 Bq cm-2, ranging from 1.1% to 20.9%. For the smaller surface area (1 cm2), detection limits were substantially higher: 1 MBq cm-2 for 99mTc and 40 Bq cm-2 for 223Ra, 177Lu, and 18F. CONCLUSION: Measurement sensitivity and accuracy depend strongly on radionuclide decay mode and contaminated area size. Small, concentrated contaminations are detectable only at sufficiently high surface activity whereas lower surface activity values become detectable when the size of the contaminated area increases. Importantly, contamination meters demonstrated optimal performance for alpha and beta emitters, which represent the most significant contamination risk in clinical practice.

Health Physics
University of Eastern Finland (FI), Kuopio University Hospital (FI)
Good health and well-being
Radioactive Decay and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon