Characterizing functional range of motion in two shoulder arthroscopy arm holders: a preliminary descriptive study

Introduction Two common arm holders utilized during shoulder arthroscopy include the Spider2 and the Trimano. The Trimano features a four-joint, manually controlled design, while the Spider2 uses a two-joint, battery-powered foot pedal system. Their differences in design may alter their range of motion (ROM) during shoulder surgery, however, functional descriptive ROM for these devices do not exist. Purpose To describe the ROM of the Trimano and Spider2 arm holders in standard arm positions for shoulder surgeries. Methods Three participants ( M = 2; age=32.3 ± 6.0years, height = 179.1 ± 16.6 cm, mass = 95.3 ± 30.7 kg), participated in this study. Three 2-D cameras captured motion in the sagittal, frontal, and transverse planes. Baseline ROM of the arm holders were captured. Subjects then completed shoulder forward flexion (FLEX), abduction (ABU) and external rotation (EXT) at baseline then with both arm holders. Dartfish360 was utilized to extract all ROM measures. Descriptive statistics were calculated for all ROM measurements. Results Arm holder baseline ROM was greater for the Trimano compared to the Spider2 across all planes. With participants, overall, the Spider2 permitted 31.13° more FLEX and 22.30° more ABU than the Trimano, while the Trimano allowed 9.93° more EXT. Conclusions The Trimano exhibited greater baseline ROM compared to the Spider2, but this did not translate when supporting the participants arms’ potentially due to the design of each device, although this was not assessed in this study. Future research should assess these arm holders in a larger population. Clinical relevance This study is the first to report functional ROM for the Trimano and Spider2 arm holders.

Authors

Institutions

Publication Details

Journal
Frontiers in Medical Technology
Published
2026-11-16
DOI
https://doi.org/10.3389/fmedt.2026.1935729
Primary Topic
Shoulder Injury and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Characterizing functional range of motion in two shoulder arthroscopy arm holders: a preliminary descriptive study

Shawn M. Gee, Patrick K. Mescher, Madison S. Mach, Michael A. Donohue et al.
3 links
Frontiers in Medical Technology
Shoulder Injury and Treatment
article

Characterizing functional range of motion in two shoulder arthroscopy arm holders: a preliminary descriptive study

Shawn M. Gee, Patrick K. Mescher, Madison S. Mach, Michael A. Donohue, Savannah F. Prosperie
article en

Abstract

Introduction Two common arm holders utilized during shoulder arthroscopy include the Spider2 and the Trimano. The Trimano features a four-joint, manually controlled design, while the Spider2 uses a two-joint, battery-powered foot pedal system. Their differences in design may alter their range of motion (ROM) during shoulder surgery, however, functional descriptive ROM for these devices do not exist. Purpose To describe the ROM of the Trimano and Spider2 arm holders in standard arm positions for shoulder surgeries. Methods Three participants ( M = 2; age=32.3 ± 6.0years, height = 179.1 ± 16.6 cm, mass = 95.3 ± 30.7 kg), participated in this study. Three 2-D cameras captured motion in the sagittal, frontal, and transverse planes. Baseline ROM of the arm holders were captured. Subjects then completed shoulder forward flexion (FLEX), abduction (ABU) and external rotation (EXT) at baseline then with both arm holders. Dartfish360 was utilized to extract all ROM measures. Descriptive statistics were calculated for all ROM measurements. Results Arm holder baseline ROM was greater for the Trimano compared to the Spider2 across all planes. With participants, overall, the Spider2 permitted 31.13° more FLEX and 22.30° more ABU than the Trimano, while the Trimano allowed 9.93° more EXT. Conclusions The Trimano exhibited greater baseline ROM compared to the Spider2, but this did not translate when supporting the participants arms’ potentially due to the design of each device, although this was not assessed in this study. Future research should assess these arm holders in a larger population. Clinical relevance This study is the first to report functional ROM for the Trimano and Spider2 arm holders.

Frontiers in Medical TechnologyVol. 8
Louisiana State University (US), Henry M. Jackson Foundation (US), Alexander T. Augusta Military Medical Center (US), United States Military Academy (US), Keller Army Community Hospital (US), Louisiana State University Health Sciences Center Shreveport (US)
Good health and well-being
Shoulder Injury and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon