In Vivo Ankle Tendon Quantification Using Ultrashort Echo Time Magnetization Transfer Imaging

OBJECTIVES: To quantitatively assess in vivo ankle tendons using less magic angle-sensitive 3-dimensional ultrashort echo time magnetization transfer (UTE MT) imaging and to compare its intersubject variability with UTE T2* mapping in healthy volunteers. MATERIALS AND METHODS: Thirty healthy volunteer ankles were examined on a 3T MRI system using a standardized clinical protocol with additional 3D UTE MT and 3D multi-echo UTE T2* sequences (acquisition times: 4:16 and 9:52 minutes, respectively). Nine major ankle tendons were semi-automatically segmented to extract quantitative UTE MTR and UTE T2* values for each tendon. Variability for each sequence was assessed using the coefficient of variation (CV), and differences in CV were evaluated primarily with a paired bootstrap of the CV ratio and secondarily with a modified signed log-likelihood ratio (MSLR) test, both with Holm correction. Within-tendon angular dependence was assessed using local tendon fiber angles relative to B0. RESULTS: Absolute quantitative values differed between tendons, with the Achilles tendon demonstrating the lowest T2* values (2.15±0.43 ms) and highest MTR values (0.44±0.08), whereas smaller positional tendons showed higher T2* and lower MTR values. UTE MTR demonstrated lower overall variability compared with UTE T2* values across the 9 assessed tendons (overall CV UTE MTR: 18.0%±4.2% vs. UTE T2*: 33.7%±8.9%, mean±SD, n=9). In the paired bootstrap analysis with Holm correction, UTE MTR showed significantly lower variability than UTE T2* in 6 of 9 tendons, with the largest differences in the extensor digitorum longus, peroneus brevis, and flexor hallucis longus tendons (T2*/MTR CV ratio, 2.74, 2.60, and 2.48; all Holm-adjusted P≤0.015); MSLR results were consistent. Between 0 degree and the magic angle, UTE T2* increased by 79% (95% CI: 64%-95%), whereas UTE MTR decreased by 9% (95% CI: 7%-11%). UTE MTR consistently demonstrated reduced dispersion and narrower value distributions compared with UTE T2*. CONCLUSIONS: In vivo ankle tendon quantification using UTE MT is feasible within clinically acceptable acquisition times and shows significantly lower intersubject variability than UTE T2* mapping in most tendons, with reduced susceptibility to the magic angle effect. UTE MT represents a promising approach for quantitative assessment of tendon structure in vivo.

Authors

Institutions

Publication Details

Journal
Investigative Radiology
Published
2026-11-16
DOI
https://doi.org/10.1097/rli.0000000000001322
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

In Vivo Ankle Tendon Quantification Using Ultrashort Echo Time Magnetization Transfer Imaging

Reto Sutter, Stefan Sommer, Georg C. Feuerriegel, Adrian A. Marth et al.
2 links
Investigative Radiology
Advanced MRI Techniques and Applications
article

In Vivo Ankle Tendon Quantification Using Ultrashort Echo Time Magnetization Transfer Imaging

Reto Sutter, Stefan Sommer, Georg C. Feuerriegel, Adrian A. Marth, Damian Mialkowskyj
article en

Abstract

OBJECTIVES: To quantitatively assess in vivo ankle tendons using less magic angle-sensitive 3-dimensional ultrashort echo time magnetization transfer (UTE MT) imaging and to compare its intersubject variability with UTE T2* mapping in healthy volunteers. MATERIALS AND METHODS: Thirty healthy volunteer ankles were examined on a 3T MRI system using a standardized clinical protocol with additional 3D UTE MT and 3D multi-echo UTE T2* sequences (acquisition times: 4:16 and 9:52 minutes, respectively). Nine major ankle tendons were semi-automatically segmented to extract quantitative UTE MTR and UTE T2* values for each tendon. Variability for each sequence was assessed using the coefficient of variation (CV), and differences in CV were evaluated primarily with a paired bootstrap of the CV ratio and secondarily with a modified signed log-likelihood ratio (MSLR) test, both with Holm correction. Within-tendon angular dependence was assessed using local tendon fiber angles relative to B0. RESULTS: Absolute quantitative values differed between tendons, with the Achilles tendon demonstrating the lowest T2* values (2.15±0.43 ms) and highest MTR values (0.44±0.08), whereas smaller positional tendons showed higher T2* and lower MTR values. UTE MTR demonstrated lower overall variability compared with UTE T2* values across the 9 assessed tendons (overall CV UTE MTR: 18.0%±4.2% vs. UTE T2*: 33.7%±8.9%, mean±SD, n=9). In the paired bootstrap analysis with Holm correction, UTE MTR showed significantly lower variability than UTE T2* in 6 of 9 tendons, with the largest differences in the extensor digitorum longus, peroneus brevis, and flexor hallucis longus tendons (T2*/MTR CV ratio, 2.74, 2.60, and 2.48; all Holm-adjusted P≤0.015); MSLR results were consistent. Between 0 degree and the magic angle, UTE T2* increased by 79% (95% CI: 64%-95%), whereas UTE MTR decreased by 9% (95% CI: 7%-11%). UTE MTR consistently demonstrated reduced dispersion and narrower value distributions compared with UTE T2*. CONCLUSIONS: In vivo ankle tendon quantification using UTE MT is feasible within clinically acceptable acquisition times and shows significantly lower intersubject variability than UTE T2* mapping in most tendons, with reduced susceptibility to the magic angle effect. UTE MT represents a promising approach for quantitative assessment of tendon structure in vivo.

Investigative Radiology
University of California, San Francisco (US), University of Zurich (CH), Universitätsklinik Balgrist (CH), Ludwig-Maximilians-Universität München (DE)
Good health and well-being
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon