Assessing Visual Field Defects in Glaucoma Using the Perceptive Scotoma Finder

Purpose: To evaluate the diagnostic performance of the Perceptive Scotoma Finder, a dynamic touchscreen-based tool, for detecting glaucomatous visual field defects and examining its structure-function associations. Methods: From the initial cohort of 143 eyes (86 participants), one eye per participant was included to avoid inter-eye correlation bias, comprising 58 glaucomatous eyes and 28 control eyes. All participants underwent monocular testing with a 30 Hz flickering dot stimulus mapped onto the Humphrey Visual Field 24-2 grid. Sectoral detection scores were compared with Humphrey Visual Field indices and retinal nerve fiber layer (RNFL) thickness. Group differences were analyzed with Mann-Whitney U tests, correlations with Spearman's rho, and diagnostic accuracy with receiver operating characteristic analysis. Test-retest reliability of the mean sector score was assessed in a subset of 26 eyes, including 18 glaucoma patients and 8 control subjects, using a two-way random-effects model for absolute agreement. Results: Perceptive Scotoma Finder detection proportions were markedly higher in glaucoma than in controls across all Garway-Heath sectors (e.g., IN: 85.7% vs 1.9%; all p < 0.001). Detection scores showed correlations with mean deviation (ρ = 0.29 to 0.50) and pattern standard deviation (ρ = 0.36 to 0.69; all p < 0.001), and with RNFL thickness in all sectors except the temporal sector (p = 0.252). Sensitivity ranged from 0.68 to 0.91 and specificity from 0.83 to 0.90 across sectors. Test-retest reliability, based on 26 eyes, was 0.76 (95% confidence interval 0.72-0.89) by ICC, and Bland-Altman analysis showed a mean bias of 0.00 with 95% limits of agreement from -0.24 to 0.24. Overall discriminative performance produced an area under the curve of 0.79 (95% Confidence interval 0.76-0.83). Conclusions: The Perceptive Scotoma Finder demonstrated reproducible, sector-specific detection of established glaucomatous visual field defects and showed moderate agreement with conventional visual field and RNFL measures. These findings support its feasibility and potential as a rapid, low-cost adjunctive tool for glaucoma assessment; however, further evaluation in early glaucoma and screening populations is required.

Authors

Institutions

Publication Details

Journal
Korean Journal of Ophthalmology
Published
2026-11-16
DOI
https://doi.org/10.3341/kjo.2026.0150
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Assessing Visual Field Defects in Glaucoma Using the Perceptive Scotoma Finder

Hyung Bin Hwang, Huma Khurshid, Yong Chan Kim, Haeyoung Lopilly Park et al.
Korean Journal of Ophthalmology
Glaucoma and retinal disorders
article

Assessing Visual Field Defects in Glaucoma Using the Perceptive Scotoma Finder

Hyung Bin Hwang, Huma Khurshid, Yong Chan Kim, Haeyoung Lopilly Park, Chan Kee Park
article en

Abstract

Purpose: To evaluate the diagnostic performance of the Perceptive Scotoma Finder, a dynamic touchscreen-based tool, for detecting glaucomatous visual field defects and examining its structure-function associations. Methods: From the initial cohort of 143 eyes (86 participants), one eye per participant was included to avoid inter-eye correlation bias, comprising 58 glaucomatous eyes and 28 control eyes. All participants underwent monocular testing with a 30 Hz flickering dot stimulus mapped onto the Humphrey Visual Field 24-2 grid. Sectoral detection scores were compared with Humphrey Visual Field indices and retinal nerve fiber layer (RNFL) thickness. Group differences were analyzed with Mann-Whitney U tests, correlations with Spearman's rho, and diagnostic accuracy with receiver operating characteristic analysis. Test-retest reliability of the mean sector score was assessed in a subset of 26 eyes, including 18 glaucoma patients and 8 control subjects, using a two-way random-effects model for absolute agreement. Results: Perceptive Scotoma Finder detection proportions were markedly higher in glaucoma than in controls across all Garway-Heath sectors (e.g., IN: 85.7% vs 1.9%; all p < 0.001). Detection scores showed correlations with mean deviation (ρ = 0.29 to 0.50) and pattern standard deviation (ρ = 0.36 to 0.69; all p < 0.001), and with RNFL thickness in all sectors except the temporal sector (p = 0.252). Sensitivity ranged from 0.68 to 0.91 and specificity from 0.83 to 0.90 across sectors. Test-retest reliability, based on 26 eyes, was 0.76 (95% confidence interval 0.72-0.89) by ICC, and Bland-Altman analysis showed a mean bias of 0.00 with 95% limits of agreement from -0.24 to 0.24. Overall discriminative performance produced an area under the curve of 0.79 (95% Confidence interval 0.76-0.83). Conclusions: The Perceptive Scotoma Finder demonstrated reproducible, sector-specific detection of established glaucomatous visual field defects and showed moderate agreement with conventional visual field and RNFL measures. These findings support its feasibility and potential as a rapid, low-cost adjunctive tool for glaucoma assessment; however, further evaluation in early glaucoma and screening populations is required.

Korean Journal of Ophthalmology
The Catholic University of Korea Incheon St. Mary's Hospital (KR), The Catholic University of Korea Seoul St. Mary's Hospital (KR), Catholic University of Korea (KR)
Good health and well-being
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon