Circular-Arc Modal Migration for Ultrasound Imaging--Part II: Simulation and Experimental Evaluation in Compact Semicircular and Convex Arrays

Circular-arc ultrasound imaging requires reconstruction that accounts for curvature and offset transmit sources across a wide field. This study evaluates how Circular-Arc Modal Migration (CAMM) changes point-response width, image-energy distribution, and low-echo target separation relative to delay-and-sum (DAS). CAMM combines angular modal analysis, local spectral remapping, and radial synthesis, with the offset phase evaluated through phase screens or approximated by scalar pullback. Guided by Part I's error analysis and point-target comparisons, paired reconstructions use scalar pullback for compact semicircular cyst simulations and phase screens for acquisition-matched C5-2v simulations, six phantom records, and two carotid views from one participant. Each DAS/CAMM pair uses identical RF. The principal observation is that lower energy fractions outside a fixed neighborhood can coexist with broader tangential responses. In C5-2v point simulations, the full-field fraction is lower at 23 of 24 locations, whereas tangential width is greater at all 24. Semicircular cyst contrast improves in most configurations, with reversed ranking at the smallest target. All eight C5-2v cyst configurations at 10--15-mm radial depth favor CAMM in the three contrast measures; deeper comparisons depend on location and metric. Measured reflectors show the same width--energy-fraction pattern, and the evaluated phantom cyst and carotid lumen regions show improved contrast. At 512 × 513 pixels and 24 transmissions, median host latency is 29.79 ms for DAS and 51.87 ms for CAMM on an RTX 4090. The tested CAMM implementation therefore provides regional gains in low-echo target separation alongside broader tangential responses and higher computational cost.

Authors

Institutions

Publication Details

Journal
bioRxiv (Cold Spring Harbor Laboratory)
Published
2026-11-16
DOI
https://doi.org/10.64898/2026.11.11.766019
Primary Topic
Ultrasound Imaging and Elastography
Type
preprint

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Circular-Arc Modal Migration for Ultrasound Imaging--Part II: Simulation and Experimental Evaluation in Compact Semicircular and Convex Arrays

Zihan Huang, Dawei Wu, Haixia Qi, Chuan Chen et al.
2 links
bioRxiv (Cold Spring Harbor Laboratory)
Ultrasound Imaging and Elastography
preprint

Circular-Arc Modal Migration for Ultrasound Imaging--Part II: Simulation and Experimental Evaluation in Compact Semicircular and Convex Arrays

Zihan Huang, Dawei Wu, Haixia Qi, Chuan Chen, Zhoufeng Liu, Chuan Chen
preprint en

Abstract

Circular-arc ultrasound imaging requires reconstruction that accounts for curvature and offset transmit sources across a wide field. This study evaluates how Circular-Arc Modal Migration (CAMM) changes point-response width, image-energy distribution, and low-echo target separation relative to delay-and-sum (DAS). CAMM combines angular modal analysis, local spectral remapping, and radial synthesis, with the offset phase evaluated through phase screens or approximated by scalar pullback. Guided by Part I's error analysis and point-target comparisons, paired reconstructions use scalar pullback for compact semicircular cyst simulations and phase screens for acquisition-matched C5-2v simulations, six phantom records, and two carotid views from one participant. Each DAS/CAMM pair uses identical RF. The principal observation is that lower energy fractions outside a fixed neighborhood can coexist with broader tangential responses. In C5-2v point simulations, the full-field fraction is lower at 23 of 24 locations, whereas tangential width is greater at all 24. Semicircular cyst contrast improves in most configurations, with reversed ranking at the smallest target. All eight C5-2v cyst configurations at 10--15-mm radial depth favor CAMM in the three contrast measures; deeper comparisons depend on location and metric. Measured reflectors show the same width--energy-fraction pattern, and the evaluated phantom cyst and carotid lumen regions show improved contrast. At 512 × 513 pixels and 24 transmissions, median host latency is 29.79 ms for DAS and 51.87 ms for CAMM on an RTX 4090. The tested CAMM implementation therefore provides regional gains in low-echo target separation alongside broader tangential responses and higher computational cost.

bioRxiv (Cold Spring Harbor Laboratory)
Second Military Medical University (CN), Southeast University (CN), Nanjing University of Aeronautics and Astronautics (CN)
National Natural Science Foundation of China, Government of Jiangsu Province, Natural Science Foundation of Jiangsu Province
Ultrasound Imaging and Elastography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon