High dose-rate response and individual correction of commercial electronic personal dosimeters for nuclear emergency radiation protection
Reliable active dosimetry is essential during nuclear and radiological emergencies, yet the response of deployed electronic personal dosimeters (EPDs) is rarely verified for every unit in the 1– 7 S v h − 1 regime. Type testing establishes model-level compliance, whereas emergency use depends on individual instruments. We characterised ten commercial silicon-diode EPD units of the same IEC 61526-certified model in a traceable 137 Cs reference field. The units showed high repeatability ( 𝐶 𝑉 < 2 % ) but unit-dependent dose-rate bias, from −9.3% to +7.7% at 7 S v h − 1 ; one unit over-responded progressively, which a single dead-time correction cannot reproduce. A device-specific second-order polynomial correction reduced the mean RMSE from 0.187 to 0 . 0 1 1 S v h − 1 and held the residual instrumental non-linearity within ± 1 . 5 % . Applying the first-session coefficients to an independent second session, and predicting intermediate dose rates not used in fitting, both kept residuals within ± 1 . 5 % , confirming temporal reproducibility and out-of-sample validity. Bare-device directional measurements showed up to ≈ 1 9 % lateral response reduction, supporting an optional conservative planning margin ( 𝑆 𝐹 ≈ 1 . 2 ). All corrections were established for a single 137 Cs spectrum and must not be extrapolated to other photon energies without re-calibration. Individual pre-deployment characterisation is a practical complement to model-level type testing for EPDs in nuclear emergency radiation protection.
Authors
- Jang-Oh Kim (ORCID: https://orcid.org/0000-0003-4705-5706)
- Gyoungjun Choi
- Ku Kang (ORCID: https://orcid.org/0000-0001-8739-218X)
- Myeongsik Shin (ORCID: https://orcid.org/0009-0005-1189-4366)
- Hyungwoo You
- Wonseok Bak (ORCID: https://orcid.org/0000-0002-9716-9085)
- Jun Yoo
Institutions
- Inje University (KR)
- Korea Atomic Energy Research Institute (KR)
Publication Details
- Journal
- Nuclear Engineering and Technology
- Published
- 2026-11-16
- DOI
- https://doi.org/10.1016/j.net.2026.104811
- Primary Topic
- Radioactive Decay and Measurement Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00