A Framework for Evaluating the Endurance Properties of Custom‐Made Osseointegrated Prostheses for Transfemoral Amputees Combining In Silico Simulations and Mechanical Testing

ABSTRACT According to the European Medical Device Regulation (MDR), new or modified implants must undergo preclinical testing to characterise their safety and performance. However, no specific standard defines how to evaluate the endurance behaviour of osseointegrated transfemoral prostheses, making the mechanical evaluation of such devices challenging. To address this limitation, the present study aimed to develop a combined experimental–numerical framework to support the endurance assessment of custom‐made osseointegrated transfemoral prostheses. The rationale for endurance testing was developed based on the loading and boundary conditions defined in the ISO 7206 standards for hip stems. A simplified stem geometry instrumented with six strain gauge rosettes was used to measure strains under controlled quasi‐static loading conditions. In parallel, a finite element (FE) model of the same experimental configuration was developed. The predictive accuracy of the FE model was assessed through direct comparison with experimentally measured strains. A sensitivity analysis was performed to evaluate the influence of uncertainties in the input parameters on FE predictions. The agreement between experimental and numerical results was excellent, with R 2 ≥ 0.998 for all gauges. The normalised root mean square error (%RMSE) was consistently below 2.7%, whereas the peak error remained below 3.9%. These findings supported the ability of the numerical methodology to reproduce the experimental loading conditions and accurately predict local strains. The numerical methodology, validated against experimental measurements using the simplified stem geometry, was subsequently applied to the OsteoCustom stem as a proof of concept. Overall, this work demonstrates the feasibility of integrating experimentally validated FE modelling into an endurance‐assessment framework for osseointegrated stems. The proposed methodology can support the identification of mechanically critical design configurations and guide subsequent experimental endurance testing of custom‐made implants.

Authors

Institutions

Publication Details

Journal
Strain
Published
2026-11-16
DOI
https://doi.org/10.1111/str.70045
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A Framework for Evaluating the Endurance Properties of Custom‐Made Osseointegrated Prostheses for Transfemoral Amputees Combining In Silico Simulations and Mechanical Testing

Giulia Galteri, Cristina Curreli, Luca Cristofolini, Marco Viceconti et al.
3 links
Strain
Orthopaedic implants and arthroplasty
article

A Framework for Evaluating the Endurance Properties of Custom‐Made Osseointegrated Prostheses for Transfemoral Amputees Combining In Silico Simulations and Mechanical Testing

Giulia Galteri, Cristina Curreli, Luca Cristofolini, Marco Viceconti, Giacomo Dozza
article en

Abstract

ABSTRACT According to the European Medical Device Regulation (MDR), new or modified implants must undergo preclinical testing to characterise their safety and performance. However, no specific standard defines how to evaluate the endurance behaviour of osseointegrated transfemoral prostheses, making the mechanical evaluation of such devices challenging. To address this limitation, the present study aimed to develop a combined experimental–numerical framework to support the endurance assessment of custom‐made osseointegrated transfemoral prostheses. The rationale for endurance testing was developed based on the loading and boundary conditions defined in the ISO 7206 standards for hip stems. A simplified stem geometry instrumented with six strain gauge rosettes was used to measure strains under controlled quasi‐static loading conditions. In parallel, a finite element (FE) model of the same experimental configuration was developed. The predictive accuracy of the FE model was assessed through direct comparison with experimentally measured strains. A sensitivity analysis was performed to evaluate the influence of uncertainties in the input parameters on FE predictions. The agreement between experimental and numerical results was excellent, with R 2 ≥ 0.998 for all gauges. The normalised root mean square error (%RMSE) was consistently below 2.7%, whereas the peak error remained below 3.9%. These findings supported the ability of the numerical methodology to reproduce the experimental loading conditions and accurately predict local strains. The numerical methodology, validated against experimental measurements using the simplified stem geometry, was subsequently applied to the OsteoCustom stem as a proof of concept. Overall, this work demonstrates the feasibility of integrating experimentally validated FE modelling into an endurance‐assessment framework for osseointegrated stems. The proposed methodology can support the identification of mechanically critical design configurations and guide subsequent experimental endurance testing of custom‐made implants.

StrainVol. 62(6)
Istituto Ortopedico Rizzoli (IT), University of Bologna (IT)
Good health and well-being, Industry, innovation and infrastructure
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon