Performance characterization of syringes used for intravitreal injections: dosing accuracy, particulate contamination, and residual volume

Abstract Background Intravitreal injections are the most frequently performed intraocular procedure worldwide. Yet the 50 µL dose is usually delivered with syringes designed neither for volumes this small nor for injection into the eye. Dosing accuracy, particulate contamination and residual volume have previously been examined in separate studies, so trade-offs between these properties within a single syringe remain unclear. We assessed all three properties in the same syringes, including models in clinical use for intravitreal injection that have not been characterised previously. Methods Six syringes were evaluated: Injekt ® -F Luer Solo 1 mL, Omnifix-F Luer Lock Solo 1 mL, Henke-Ject eco 1 mL (three-part), Zero Residual 0.2 mL, BD Micro-Fine 1 mL, and Omnican 50 U-100 insulin syringe. Dosing accuracy and precision for a 50 µL target volume was assessed with 23 specialist staff. The dosing data were analysed descriptively, and operator-dependent variability was quantified with a linear mixed model. Residual volume was determined according to DIN EN ISO 7886-1. Particulate contamination was quantified by light obscuration after 7-day incubation of water, with and without flicking of the syringe during filling. Particle counts were analyzed descriptively. Results All syringes tended to overdose, with a median delivered volume of 55.1 µL, and only 45.4% of the measurements were within 45–55 µL. In the mixed model, 22% of the variance of the delivered volume was attributable to between-operator differences, and no syringe type differed significantly in mean delivered volume. Among the syringes measured without a needle, the mean residual volume ranged from 0.18 µL to 23.1 µL; the two syringes with integrated needles showed 0.59 µL and 1.77 µL including the needle dead space. All syringes released particulate matter, and in this exploratory study the particle count was higher after flicking in five of the six syringe types. Conclusions None of the evaluated syringes performed optimally across all properties assessed. The Zero Residual 0.2 mL showed the lowest residual volume of the syringes measured without a needle and low particle concentrations without flicking, but delivered individual doses well outside the target range and showed a ten-fold rise in particle concentration after flicking. Repurposed syringes delivered volumes of up to 98 µL.

Authors

Institutions

Publication Details

Journal
International Journal of Retina and Vitreous
Published
2026-11-16
DOI
https://doi.org/10.1186/s40942-026-00967-3
Primary Topic
Intravenous Infusion Technology and Safety
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Performance characterization of syringes used for intravitreal injections: dosing accuracy, particulate contamination, and residual volume

Anne Rübsam, Michael Lommel, Felix Hehnen, Tim Bierewirtz et al.
1 links
International Journal of Retina and Vitreous
Intravenous Infusion Technology and Safety
article

Performance characterization of syringes used for intravitreal injections: dosing accuracy, particulate contamination, and residual volume

Anne Rübsam, Michael Lommel, Felix Hehnen, Tim Bierewirtz, Paul Friedrich Geus, Elias Köhne, Ulrich Kertzscher, Susan Henniger, Fabian Lohmüller
article en

Abstract

Abstract Background Intravitreal injections are the most frequently performed intraocular procedure worldwide. Yet the 50 µL dose is usually delivered with syringes designed neither for volumes this small nor for injection into the eye. Dosing accuracy, particulate contamination and residual volume have previously been examined in separate studies, so trade-offs between these properties within a single syringe remain unclear. We assessed all three properties in the same syringes, including models in clinical use for intravitreal injection that have not been characterised previously. Methods Six syringes were evaluated: Injekt ® -F Luer Solo 1 mL, Omnifix-F Luer Lock Solo 1 mL, Henke-Ject eco 1 mL (three-part), Zero Residual 0.2 mL, BD Micro-Fine 1 mL, and Omnican 50 U-100 insulin syringe. Dosing accuracy and precision for a 50 µL target volume was assessed with 23 specialist staff. The dosing data were analysed descriptively, and operator-dependent variability was quantified with a linear mixed model. Residual volume was determined according to DIN EN ISO 7886-1. Particulate contamination was quantified by light obscuration after 7-day incubation of water, with and without flicking of the syringe during filling. Particle counts were analyzed descriptively. Results All syringes tended to overdose, with a median delivered volume of 55.1 µL, and only 45.4% of the measurements were within 45–55 µL. In the mixed model, 22% of the variance of the delivered volume was attributable to between-operator differences, and no syringe type differed significantly in mean delivered volume. Among the syringes measured without a needle, the mean residual volume ranged from 0.18 µL to 23.1 µL; the two syringes with integrated needles showed 0.59 µL and 1.77 µL including the needle dead space. All syringes released particulate matter, and in this exploratory study the particle count was higher after flicking in five of the six syringe types. Conclusions None of the evaluated syringes performed optimally across all properties assessed. The Zero Residual 0.2 mL showed the lowest residual volume of the syringes measured without a needle and low particle concentrations without flicking, but delivered individual doses well outside the target range and showed a ten-fold rise in particle concentration after flicking. Repurposed syringes delivered volumes of up to 98 µL.

International Journal of Retina and Vitreous
Humboldt-Universität zu Berlin (DE), Deutsches Herzzentrum der Charité (DE), German Centre for Cardiovascular Research (DE), Berlin Institute of Health at Charité - Universitätsmedizin Berlin (DE), Institute of Science Tokyo (JP), Laboratory for Future Interdisciplinary Research of Science and Technology (JP), Freie Universität Berlin (DE), Charité - Universitätsmedizin Berlin (DE)
Charité – Universitätsmedizin Berlin
Good health and well-being
Intravenous Infusion Technology and Safety
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.