Finite element analysis of different subtrochanteric osteotomy techniques in total hip arthroplasty for high hip dislocation
Aims Subtrochanteric shortening osteotomy in total hip arthroplasty (THA) for developmental dysplasia of the hip (DDH) has variable outcomes. This study evaluated the initial stability of seven different osteotomy techniques. Methods A 3D finite element model of a femur with a stem was created. Seven models simulating osteotomies with different angles (0°, 45°, 90°, -90°, 120°, -120°, z-shaped) were analyzed. Outcomes included range of motion (ROM), bone/implant stress, and osteotomy surface separation. Results The 0° model had the greatest ROM (lowest stiffness). The z-shaped and -90° designs provided the best rotational stability. Analysis of the centre of rotation showed single-symmetric models (45°, z-shaped) shifted laterally, suggesting a ‘windshield wiper’ effect, while double-symmetric models (90°, -90°, 120°, -120°) stayed aligned with the prosthesis axis. Stress patterns differed: single-symmetric models concentrated stress on the lesser trochanter, while double-symmetric models stressed the greater trochanter, with a notable ring in the -90° design. Implant stresses were similar, except in the horizontal model. The -90° and 90° models had the least osteotomy separation, unlike the 45° model. Conclusion No single osteotomy technique outperformed others in all metrics. The z-shaped and -90° designs offered the best rotational stability; the -90° and 90° designs minimized axial separation; the 45° model showed the worst ‘windshield wiper’ effect. The 120° technique provided a balanced profile with moderate rotational stability, acceptable separation, and less pronounced stress rings, representing a reasonable compromise. The choice should be guided by clinical priorities. Cite this article: Bone Joint Res 2026;15(11):1364–1373.
Authors
- huifa xu
- Long Bi (ORCID: https://orcid.org/0000-0002-9854-9338)
- Bingchen Xia
- 韩一生
- Weilong Diwu
- Zhong Wang (ORCID: https://orcid.org/0000-0002-0975-6707)
- Yuanqi Hong
- Min Yang (ORCID: https://orcid.org/0009-0007-1556-2830)
- Qi Wu (ORCID: https://orcid.org/0009-0000-0547-7914)
Institutions
- Dalian University of Technology (CN)
- First Affiliated Hospital of Jiangxi Medical College (CN)
- Xi'an Honghui Hospital (CN)
- Jiangxi Provincial People's Hospital (CN)
- Xijing Hospital (CN)
- Shaanxi University of Science and Technology (CN)
- Xi'an Jiaotong University (CN)
- Northeastern University (CN)
- Air Force Medical University (CN)
Publication Details
- Journal
- Bone and Joint Research
- Published
- 2026-11-16
- DOI
- https://doi.org/10.1302/2046-3758.1511.bjr-2026-0216.r1
- Primary Topic
- Orthopaedic implants and arthroplasty
- Type
- article
- Field-Weighted Citation Impact
- 0.00