Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat Eyes

OBJECTIVE: To validate the accuracy of the iFalcon V700 rebound tonometer in measuring intraocular pressure (IOP) in enucleated normal rat eyes. ANIMALS STUDIED: Ten freshly enucleated normal rat eyes were tested within 12 h postmortem. PROCEDURES: Enucleated rat eyes were fixed to the front of the device. IOP measurements started at 5 mmHg and were gradually increased to 95 mmHg by mechanical pressure from the vitreous side toward the anterior chamber. The iFalcon V700 rebound tonometer measured central corneal IOP in rodent mode, and results were compared with the iCare TonoLab using linear regression analysis and Bland-Altman plots. RESULTS: = 0.98). Bland-Altman analysis showed a mean bias of -1.56 mmHg, with 95% limits of agreement from -6.71 to +3.59 mmHg, and slight underestimation in the moderate IOP range. In the 5-10 mmHg range, multiple attempts were needed to get valid readings with the iFalcon V700. CONCLUSIONS: The iFalcon V700 enables accurate central-corneal IOP measurement (5-95 mmHg) in enucleated normal rat eyes and shows good agreement with iCare TonoLab; repeated measurements at identical pressure levels improve precision. This study has limitations: inherent errors from iCare TonoLab cannot be excluded, and mechanical compression precludes testing at extremely low IOP (0-5 mmHg). Despite its promising potential for rodent tonometry, further systematic in vivo validation is required prior to its widespread, reliable use in rat ophthalmic research.

Authors

Institutions

Publication Details

Journal
Veterinary Ophthalmology
Published
2026-11-13
DOI
https://doi.org/10.1111/vop.70300
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat Eyes

Fu Ji, Yunhao Su, Lihui Yang, Haobo Zhao
Veterinary Ophthalmology
Glaucoma and retinal disorders
article

Inter‐Device Consistency Between iFalcon and iCare in Mechanically Pressurized Ex Vivo Rat Eyes

Fu Ji, Yunhao Su, Lihui Yang, Haobo Zhao
article en

Abstract

OBJECTIVE: To validate the accuracy of the iFalcon V700 rebound tonometer in measuring intraocular pressure (IOP) in enucleated normal rat eyes. ANIMALS STUDIED: Ten freshly enucleated normal rat eyes were tested within 12 h postmortem. PROCEDURES: Enucleated rat eyes were fixed to the front of the device. IOP measurements started at 5 mmHg and were gradually increased to 95 mmHg by mechanical pressure from the vitreous side toward the anterior chamber. The iFalcon V700 rebound tonometer measured central corneal IOP in rodent mode, and results were compared with the iCare TonoLab using linear regression analysis and Bland-Altman plots. RESULTS: = 0.98). Bland-Altman analysis showed a mean bias of -1.56 mmHg, with 95% limits of agreement from -6.71 to +3.59 mmHg, and slight underestimation in the moderate IOP range. In the 5-10 mmHg range, multiple attempts were needed to get valid readings with the iFalcon V700. CONCLUSIONS: The iFalcon V700 enables accurate central-corneal IOP measurement (5-95 mmHg) in enucleated normal rat eyes and shows good agreement with iCare TonoLab; repeated measurements at identical pressure levels improve precision. This study has limitations: inherent errors from iCare TonoLab cannot be excluded, and mechanical compression precludes testing at extremely low IOP (0-5 mmHg). Despite its promising potential for rodent tonometry, further systematic in vivo validation is required prior to its widespread, reliable use in rat ophthalmic research.

Veterinary OphthalmologyVol. 30(1)
Zhejiang Sci-Tech University (CN)
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon