A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps

ABSTRACT Objective Variability of 3D‐printed rotodynamic blood pumps (RBP) indicators is a crucial factor in haemocompatibility evaluation. Thus, the aim is to investigate the reproducibility of a 3D printed‐based framework for further use on systematic haemocompatible studies when changing key impeller parameters in RBPs. Methods Reproducibility was assessed in terms of standard deviation (SD), of the normalized index of haemolysis (NIH) and generated pressure head between 3 RBPs. A conventionally manufactured benchmark RBP, one 3D printed (3DP) exact replica, (), and one 3DP with a modification () ( n = 7). In vitro haemolysis experiments were carried out. Newly 3DP impellers were used for each experiment to study the manufacturing reproducibility. Moreover, pressure head‐flow rate curve analysis and computational fluid dynamic (CFD) simulations were performed to support the findings. Results The mean NIH ± SD in mg/100 L for the 3 RBPs was 4.25 ± 1.95 for the benchmark, 6.40 ± 2.68 for the exact replica, and 7.75 ± 1.87 for the modified version. Only a significant difference ( p < 0.05) between the benchmark and modified version was found. Conversely, no strong statistical differences were found between the 3DP RBPs. Furthermore, SD values of pressure heads were similar between all the groups. The resulting damage index for the 3DP geometries was 2.37 (·10 −4 ) and 2.32 (·10 −4 ) for the replica and modified version respectively. Suggesting a probable small effect when variating in free stream pumps. Conclusions Same NIH and pressure head SD range has shown the viability of the 3DP experimental framework, proposing an economic and rapid path for systematic haemocompatibility studies of parameters in RBPs.

Authors

Institutions

Publication Details

Journal
Artificial Organs
Published
2026-11-13
DOI
https://doi.org/10.1111/aor.70285
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps

Benjamin Torner, Marcus Granegger, Aldo J. Suria, Niccolò Crescenti et al.
Artificial Organs
Mechanical Circulatory Support Devices
article

A 3D Print‐Based Experimental Framework for the Assessment of Haemocompatibility in Rotodynamic Blood Pumps

Benjamin Torner, Marcus Granegger, Aldo J. Suria, Niccolò Crescenti, Daniel Zimpfer
article en

Abstract

ABSTRACT Objective Variability of 3D‐printed rotodynamic blood pumps (RBP) indicators is a crucial factor in haemocompatibility evaluation. Thus, the aim is to investigate the reproducibility of a 3D printed‐based framework for further use on systematic haemocompatible studies when changing key impeller parameters in RBPs. Methods Reproducibility was assessed in terms of standard deviation (SD), of the normalized index of haemolysis (NIH) and generated pressure head between 3 RBPs. A conventionally manufactured benchmark RBP, one 3D printed (3DP) exact replica, (), and one 3DP with a modification () ( n = 7). In vitro haemolysis experiments were carried out. Newly 3DP impellers were used for each experiment to study the manufacturing reproducibility. Moreover, pressure head‐flow rate curve analysis and computational fluid dynamic (CFD) simulations were performed to support the findings. Results The mean NIH ± SD in mg/100 L for the 3 RBPs was 4.25 ± 1.95 for the benchmark, 6.40 ± 2.68 for the exact replica, and 7.75 ± 1.87 for the modified version. Only a significant difference ( p < 0.05) between the benchmark and modified version was found. Conversely, no strong statistical differences were found between the 3DP RBPs. Furthermore, SD values of pressure heads were similar between all the groups. The resulting damage index for the 3DP geometries was 2.37 (·10 −4 ) and 2.32 (·10 −4 ) for the replica and modified version respectively. Suggesting a probable small effect when variating in free stream pumps. Conclusions Same NIH and pressure head SD range has shown the viability of the 3DP experimental framework, proposing an economic and rapid path for systematic haemocompatibility studies of parameters in RBPs.

Artificial Organs
Medical University of Vienna (AT)
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon