Empirical and semi-empirical methods to determine the kV exponent (n value) for diagnostic X-ray machines.
The air kerma, and beam quality measurements were performed on 43 diagnostic X-ray machines (25 fixed and 18 mobile) that met acceptance criteria for tube potential accuracy, half-value layer, and tube output. Measurements were conducted at tube potentials from 40 kV to 120 kV using a calibrated solid-state dosimetry system at a fixed focus-detector distance of 1 m, normalized output (mGy/mAs) and half-value layer were determined at each setting of the tube potential. Two methods (method 1 - empirical and method 2 - semi-empirical) were used to derive the exponent value. In method 1, a power-law fit of tube output versus tube potential yielded an average exponent of $n$ ≈ 2.334 ± 0.118 based on the group of X-ray machines with high-power frequency generators, consistent with values reported for diagnostic systems with standard filtration [1, 2]. In method 2, the $n$ was derived as a function of tube potential by incorporating the measured variation of half-value layer with tube potential, explicitly accounting for effective filtration. This approach showed that $n$ decreases from ~2.69 at 40 kV to 2.09 at 120 kV, demonstrating greater output sensitivity to tube potential changes at lower tube potentials. Therefore, the study introduces a semi-empirical method for estimating the exponent ($n$ value) that incorporates measured half-value layer variation to account for effective beam hardening, which is neglected in conventional empirical models. This provides a practical tool for characterizing the dependence of X-ray tube output on tube protocol optimization and equipment performance assessment, applicable for quality control in diagnostic X-ray machines.
Authors
- Wellen Rukundo (ORCID: https://orcid.org/0000-0001-7705-5518)
- Natharius Nimbashabira (ORCID: https://orcid.org/0000-0002-6369-8931)
- Christine Edna Emong
- Samuel Wattisa Maberi
Institutions
- Makerere University (UG)
Publication Details
- Journal
- PubMed
- Published
- 2026-11-13
- DOI
- https://doi.org/10.1093/rpd/ncag131
- Primary Topic
- Radiation Dose and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00