Raman spectroscopy as a method for purity control of fibrinogen-containing material in the production of biomedical products

Aim– to assess the possibility of Raman spectroscopy combined with chemometric methods for non-destructive control of the purification degree of fibrinogen-containing material at successive stages of a patented technology for obtaining the compound. Material and methods. Samples at three technological stages (native precipitate, crude extract, purified product) and a commercial reference preparation were studied. Raman spectroscopy was used followed by data processing using principal component analysis (PCA) and linear discriminant analysis (LDA) to classify spectral profiles.. Results. It was shown that Raman spectra change regularly along the technological chain; the key contribution to separation is made by changes in the amide regions (Amide I and Amide III) and the phenylalanine band, reflecting variations in protein composition and conformational state. The classification accuracy of purification stages was 95%. Spectral differences between laboratory samples and the commercial preparation were established, due to the degree of purification, homogeneity of the protein fraction, and the presence of auxiliary components in the dosage form. Conclusion. Raman spectroscopy combined with chemometrics is a promising rapid non-destructive tool for industrial spectral monitoring of the purification degree of fibrinogen-containing material; however, for implementation in real technological control, an expansion of the sample set, consideration of batch-to-batch variability, and independent validation of the classification model on new batches are required.

Authors

Institutions

Publication Details

Journal
Aspirantskiy Vestnik Povolzhiya
Published
2026-11-13
DOI
https://doi.org/10.35693/avp716630
Primary Topic
Spectroscopy Techniques in Biomedical and Chemical Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Raman spectroscopy as a method for purity control of fibrinogen-containing material in the production of biomedical products

L. T. Volova, Denis G. Alekseev, Pavel E. Timchenko, Elena V. Timchenko et al.
1 links
Aspirantskiy Vestnik Povolzhiya
Spectroscopy Techniques in Biomedical and Chemical Research
article

Raman spectroscopy as a method for purity control of fibrinogen-containing material in the production of biomedical products

L. T. Volova, Denis G. Alekseev, Pavel E. Timchenko, Elena V. Timchenko, Nikolay A. Ryabov, Olga A. Karyakina, Polina I. Pyataeva
article en

Abstract

Aim– to assess the possibility of Raman spectroscopy combined with chemometric methods for non-destructive control of the purification degree of fibrinogen-containing material at successive stages of a patented technology for obtaining the compound. Material and methods. Samples at three technological stages (native precipitate, crude extract, purified product) and a commercial reference preparation were studied. Raman spectroscopy was used followed by data processing using principal component analysis (PCA) and linear discriminant analysis (LDA) to classify spectral profiles.. Results. It was shown that Raman spectra change regularly along the technological chain; the key contribution to separation is made by changes in the amide regions (Amide I and Amide III) and the phenylalanine band, reflecting variations in protein composition and conformational state. The classification accuracy of purification stages was 95%. Spectral differences between laboratory samples and the commercial preparation were established, due to the degree of purification, homogeneity of the protein fraction, and the presence of auxiliary components in the dosage form. Conclusion. Raman spectroscopy combined with chemometrics is a promising rapid non-destructive tool for industrial spectral monitoring of the purification degree of fibrinogen-containing material; however, for implementation in real technological control, an expansion of the sample set, consideration of batch-to-batch variability, and independent validation of the classification model on new batches are required.

Aspirantskiy Vestnik Povolzhiya
Samara State Medical University (RU)
Good health and well-being
Spectroscopy Techniques in Biomedical and Chemical Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon