Fully automated, fast, global noise level calculation for abdomino-pelvic adult CT images

PURPOSE: The Global Noise Level (GNL) is an established objective CT noise metric evaluated directly on patient clinical images. A free, open-source, customisable tool to acceptance-test and validate commercial GNL software applications, monitor patient image noise, optimise protocols and which permits flexible input parameters for research purposes is lacking. This working developed such a tool and investigated the impact of various user-supplied input parameters on the GNL. METHOD: The tool was developed in Python and followed standard published methods; it was validated against manual calculation. The impact of user-supplied input parameters on GNL values, namely the upper/lower CT-number threshold-fences, kernel size and shape (square and novel disc-shape) and histogram bin-widths, was studied by applying the tool to two abdomino-pelvic multi-centric, multi-system and multi-phasic image datasets (n = 357). RESULTS: The tool accurately calculated GNL (Bland-Altman plot of automatic vs manual calculation showed bias = 0 and standard deviation = 0) with a processing rate of 14 slices/s. GNL values changed significantly as a function of threshold-fence. GNL values increased monotonically with kernel size for both square and disc kernels with the latter consistently giving lower values. GNL values also increased with bin-width. CONCLUSIONS: A free, fully-automated flexible tool for GNL calculation for adult abdomino-pelvic CT was developed. Since GNL values are dependent on choice of input parameters, these should therefore be reported alongside GNL values. Recommended parameters are threshold fences of - 300:0HU for adipose and 0:170HU for soft tissue (higher when contrast is present), 3 mm radius disc-kernel, 1.0HU bin-width.

Authors

Institutions

Publication Details

Journal
Physica Medica
Published
2026-11-13
DOI
https://doi.org/10.1016/j.ejmp.2026.107237
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Fully automated, fast, global noise level calculation for abdomino-pelvic adult CT images

Gianluca Valentino, Hilde Bosmans, Carmel J. Caruana, Melvin D’Anastasi et al.
2 links
Physica Medica
Radiation Dose and Imaging
article

Fully automated, fast, global noise level calculation for abdomino-pelvic adult CT images

Gianluca Valentino, Hilde Bosmans, Carmel J. Caruana, Melvin D’Anastasi, Eric Pace, Kelvin Cortis
article en

Abstract

PURPOSE: The Global Noise Level (GNL) is an established objective CT noise metric evaluated directly on patient clinical images. A free, open-source, customisable tool to acceptance-test and validate commercial GNL software applications, monitor patient image noise, optimise protocols and which permits flexible input parameters for research purposes is lacking. This working developed such a tool and investigated the impact of various user-supplied input parameters on the GNL. METHOD: The tool was developed in Python and followed standard published methods; it was validated against manual calculation. The impact of user-supplied input parameters on GNL values, namely the upper/lower CT-number threshold-fences, kernel size and shape (square and novel disc-shape) and histogram bin-widths, was studied by applying the tool to two abdomino-pelvic multi-centric, multi-system and multi-phasic image datasets (n = 357). RESULTS: The tool accurately calculated GNL (Bland-Altman plot of automatic vs manual calculation showed bias = 0 and standard deviation = 0) with a processing rate of 14 slices/s. GNL values changed significantly as a function of threshold-fence. GNL values increased monotonically with kernel size for both square and disc kernels with the latter consistently giving lower values. GNL values also increased with bin-width. CONCLUSIONS: A free, fully-automated flexible tool for GNL calculation for adult abdomino-pelvic CT was developed. Since GNL values are dependent on choice of input parameters, these should therefore be reported alongside GNL values. Recommended parameters are threshold fences of - 300:0HU for adipose and 0:170HU for soft tissue (higher when contrast is present), 3 mm radius disc-kernel, 1.0HU bin-width.

Physica MedicaVol. 152
University of Malta (MT), Mater Dei Hospital (MT), KU Leuven (BE)
Good health and well-being
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon