Clinical feasibility of patient-specific head and neck VMAT QA in high-volume practice: workflow efficiency analysis and gamma-based agreement between cylindrical 3D diode detector array and portal dosimetry

Background: This retrospective study examines the workflow efficiency and gamma analysis outcomes of ArcCHECK TM (Sun Nuclear Corporation, Melbourne, FL, USA) and EPIbeam TM (DOSIsoft, Cachan, France) portal dosimetry for verifying head and neck volumetric modulated arc therapy (VMAT) treatment plans, with an emphasis on their suitability for high-volume radiotherapy facilities. Materials and methods: Thirty-two head and neck VMAT plans were retrospectively analysed. The Elekta Versa HDTM (Elekta, Stockholm, Sweden) and Monaco TM treatment planning system (TPS) were used to generate all plans. Workflow times were documented, and gamma analysis was carried out using multiple criteria (3%/3 mm to 1%/1 mm). Statistical comparisons included mean differences, correlation coefficients and paired t-tests. Results: Both EPIbeam TM and ArcCHECK TM performed well under the 3%/3 mm and 3%/2 mm criteria. However, EPIbeam TM portal dosimetry required less setup time, supporting improved throughput in high-volume settings. Higher mean gamma pass rates were observed with EPIbeam TM in both patient groups. For the 3%/3 mm and 3%/2 mm criteria, respectively, ArcCHECK TM achieved 98.32% ± 1.03% and 96.33% ± 2.11%, whereas EPIbeam TM achieved 99.93% ± 0.09% and 99.70% ± 0.40%. The differences were statistically significant (p < 0.01) and the low correlation coefficients indicated that the two systems respond differently to complex head and neck plans. Conclusions : Utilising both systems at different stages, depending on the requirements and device-specific tolerances, can strengthen the overall QA process and provide an added layer of confidence.

Authors

Institutions

Publication Details

Journal
Reports of Practical Oncology & Radiotherapy
Published
2026-11-13
DOI
https://doi.org/10.5603/rpor.115425
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Clinical feasibility of patient-specific head and neck VMAT QA in high-volume practice: workflow efficiency analysis and gamma-based agreement between cylindrical 3D diode detector array and portal dosimetry

David Khanna, Bharath Pandu, Maniyan Paramasivam, Venkatraman Pitchaikannu et al.
3 links
Reports of Practical Oncology & Radiotherapy
Advanced Radiotherapy Techniques
article

Clinical feasibility of patient-specific head and neck VMAT QA in high-volume practice: workflow efficiency analysis and gamma-based agreement between cylindrical 3D diode detector array and portal dosimetry

David Khanna, Bharath Pandu, Maniyan Paramasivam, Venkatraman Pitchaikannu, Venugopal Sundaram, Vignesh Dharuman, Sureshbabu Pandu
article en

Abstract

Background: This retrospective study examines the workflow efficiency and gamma analysis outcomes of ArcCHECK TM (Sun Nuclear Corporation, Melbourne, FL, USA) and EPIbeam TM (DOSIsoft, Cachan, France) portal dosimetry for verifying head and neck volumetric modulated arc therapy (VMAT) treatment plans, with an emphasis on their suitability for high-volume radiotherapy facilities. Materials and methods: Thirty-two head and neck VMAT plans were retrospectively analysed. The Elekta Versa HDTM (Elekta, Stockholm, Sweden) and Monaco TM treatment planning system (TPS) were used to generate all plans. Workflow times were documented, and gamma analysis was carried out using multiple criteria (3%/3 mm to 1%/1 mm). Statistical comparisons included mean differences, correlation coefficients and paired t-tests. Results: Both EPIbeam TM and ArcCHECK TM performed well under the 3%/3 mm and 3%/2 mm criteria. However, EPIbeam TM portal dosimetry required less setup time, supporting improved throughput in high-volume settings. Higher mean gamma pass rates were observed with EPIbeam TM in both patient groups. For the 3%/3 mm and 3%/2 mm criteria, respectively, ArcCHECK TM achieved 98.32% ± 1.03% and 96.33% ± 2.11%, whereas EPIbeam TM achieved 99.93% ± 0.09% and 99.70% ± 0.40%. The differences were statistically significant (p < 0.01) and the low correlation coefficients indicated that the two systems respond differently to complex head and neck plans. Conclusions : Utilising both systems at different stages, depending on the requirements and device-specific tolerances, can strengthen the overall QA process and provide an added layer of confidence.

Reports of Practical Oncology & Radiotherapy
Karunya University (IN), Bangalore Baptist Hospital (IN), Mangalore Institute of Oncology (IN), Zydus Lifesciences Limited (India) (IN), All India Institute of Medical Sciences (IN)
Good health and well-being
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon