Battery-free ingestible electronic capsule with programmable high-output soft actuation for active gastrointestinal interventions

Ingestible capsules have garnered considerable attention due to their high gastrointestinal (GI) transitability and minimally invasive nature. However, constrained by pill-sized dimensions (approximately 2 to 4 cubic centimeters), current ingestible capsules face persistent challenges in achieving both controllable high mechanical output and sufficient onboard power delivery, limiting their capacity for effective in vivo therapeutic interventions. Here, we present a wirelessly powered ingestible electronic capsule that integrates programmable soft phase-change actuation within a rigid-soft architecture to enable programmable, high-output mechanical operations. Leveraging bidirectional thermoelectric regulation and guided rigid-soft coupling, the capsule delivers a peak force of 3.8 newtons and a volumetric stroke of ∼250 cubic millimeters within a confined geometry based on the phase-change actuation. To power this actuation, we implemented a compact wireless power transfer system that achieved a maximum load-received power per unit receiver-coil volume of 9.15 watts per cubic centimeter at a 5-centimeter transfer distance under the tested conditions. The structural adaptability of the soft actuator allows modular integration of functional end-effectors, supporting high-output tissue acquisition, site-directed delivery of model liquid or gel payloads, and active biofluid sampling under magnetic navigation. In vivo experiments in beagles demonstrated the feasibility of mechanical interventions in physiological environments. By overcoming the long-standing power and actuation constraints of capsule architecture, this work establishes a physical framework for high-output ingestible robotics and advances the development of minimally invasive robotic platforms for complex GI interventions.

Authors

Institutions

Publication Details

Journal
Science Advances
Published
2026-11-13
DOI
https://doi.org/10.1126/sciadv.aeh3758
Primary Topic
Soft Robotics and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Battery-free ingestible electronic capsule with programmable high-output soft actuation for active gastrointestinal interventions

Hubiao Fang, Hanjie Wang, Zhuan Liao, Jiaqi Zhu et al.
3 links
Science Advances
Soft Robotics and Applications
article

Battery-free ingestible electronic capsule with programmable high-output soft actuation for active gastrointestinal interventions

Hubiao Fang, Hanjie Wang, Zhuan Liao, Jiaqi Zhu, Qing Cao, Jie Hu, Wei Tang, Li Zhang, Wei Zhang, Chen He, Dong Han, Qifeng Zhu, Ran Xie, Kexin Zhang, Huayong Yang, Xi Jiang
article en

Abstract

Ingestible capsules have garnered considerable attention due to their high gastrointestinal (GI) transitability and minimally invasive nature. However, constrained by pill-sized dimensions (approximately 2 to 4 cubic centimeters), current ingestible capsules face persistent challenges in achieving both controllable high mechanical output and sufficient onboard power delivery, limiting their capacity for effective in vivo therapeutic interventions. Here, we present a wirelessly powered ingestible electronic capsule that integrates programmable soft phase-change actuation within a rigid-soft architecture to enable programmable, high-output mechanical operations. Leveraging bidirectional thermoelectric regulation and guided rigid-soft coupling, the capsule delivers a peak force of 3.8 newtons and a volumetric stroke of ∼250 cubic millimeters within a confined geometry based on the phase-change actuation. To power this actuation, we implemented a compact wireless power transfer system that achieved a maximum load-received power per unit receiver-coil volume of 9.15 watts per cubic centimeter at a 5-centimeter transfer distance under the tested conditions. The structural adaptability of the soft actuator allows modular integration of functional end-effectors, supporting high-output tissue acquisition, site-directed delivery of model liquid or gel payloads, and active biofluid sampling under magnetic navigation. In vivo experiments in beagles demonstrated the feasibility of mechanical interventions in physiological environments. By overcoming the long-standing power and actuation constraints of capsule architecture, this work establishes a physical framework for high-output ingestible robotics and advances the development of minimally invasive robotic platforms for complex GI interventions.

Science AdvancesVol. 12(46)
Chinese University of Hong Kong (HK), Second Military Medical University (CN), Hainan University (CN), National Clinical Research Center for Digestive Diseases (CN), Changhai Hospital (CN), State Key Laboratory Fluid Power and Mechatronic Systems (CN), Zhejiang University (CN)
National Natural Science Foundation of China
Good health and well-being, Industry, innovation and infrastructure
Soft Robotics and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon