Fabrication and Mechanical Evaluation of Co–28Cr–6Mo–0.11N Alloy Spinal Correction Rods with a Prescribed Spinal Contour by Electron Beam Powder Bed Fusion

In spinal correction surgery, metallic rods are commonly bent intraoperatively by surgeons to conform to the patient-specific spinal curvature and are then fixed with pedicle screws. However, intraoperative rod contouring can introduce surface scratches and notches, which may reduce fatigue strength and act as initiation sites for fracture. In this study, spinal correction rods with a prescribed spinal contour were fabricated from a Co–28Cr–6Mo–0.11N alloy (hereafter referred to as the CCMN alloy) by electron beam powder bed fusion (PBF-EB) to reduce dependence on intraoperative bending. Their compressive response, cyclic compression behavior, and fracture characteristics were experimentally evaluated, and their deformation behavior was further examined by finite element analysis (FEA). The CCMN rods were subjected to one, two, or three cycles of combined aging and reverse transformation heat treatment, denoted RT, 2RT, and 3RT, respectively. Under a compressive load of 1000 N, the displacement decreased with an increasing number of reverse transformation cycles, and the 3RT specimen showed the highest apparent resistance to deformation. The loads required to produce a displacement of 5 mm were 812, 843, 865, and 896 N for the as-built, RT, 2RT, and 3RT rods, respectively; thus, the 3RT rod exhibited an increase of 84 N relative to the as-built rod. Although all specimens were fabricated using the same CAD geometry and adjusted to the same nominal dimensional profile, small specimen-specific geometric deviations could not be independently quantified. Therefore, the observed differences in load–displacement response cannot be attributed exclusively to heat-treatment-induced microstructural changes. Furthermore, increasing the maximum diameter of the middle region of the variable-diameter rod from 7 to 8 mm reduced the displacement under 1000 N from 3.3 to 2.7 mm. In a single-specimen cyclic compression screening test at 0–300 N, 1 Hz, and 1 × 106 cycles, no apparent macroscopic residual deformation or fracture was observed. This result represents runout under the specified loading condition and does not establish a fatigue limit. These results support the feasibility of using PBF-EB to fabricate CCMN rods with prescribed curved and variable-diameter geometries and demonstrate a stable mechanical response under the specific static and cyclic loading conditions examined in this study.

Authors

Institutions

Publication Details

Journal
Alloys
Published
2026-11-13
DOI
https://doi.org/10.3390/alloys5040033
Primary Topic
Additive Manufacturing Materials and Processes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Fabrication and Mechanical Evaluation of Co–28Cr–6Mo–0.11N Alloy Spinal Correction Rods with a Prescribed Spinal Contour by Electron Beam Powder Bed Fusion

Hao Wang, Akihiko Chiba, Qiang Li, Hao Sun
1 links
Alloys
Additive Manufacturing Materials and Processes
article

Fabrication and Mechanical Evaluation of Co–28Cr–6Mo–0.11N Alloy Spinal Correction Rods with a Prescribed Spinal Contour by Electron Beam Powder Bed Fusion

Hao Wang, Akihiko Chiba, Qiang Li, Hao Sun
article en

Abstract

In spinal correction surgery, metallic rods are commonly bent intraoperatively by surgeons to conform to the patient-specific spinal curvature and are then fixed with pedicle screws. However, intraoperative rod contouring can introduce surface scratches and notches, which may reduce fatigue strength and act as initiation sites for fracture. In this study, spinal correction rods with a prescribed spinal contour were fabricated from a Co–28Cr–6Mo–0.11N alloy (hereafter referred to as the CCMN alloy) by electron beam powder bed fusion (PBF-EB) to reduce dependence on intraoperative bending. Their compressive response, cyclic compression behavior, and fracture characteristics were experimentally evaluated, and their deformation behavior was further examined by finite element analysis (FEA). The CCMN rods were subjected to one, two, or three cycles of combined aging and reverse transformation heat treatment, denoted RT, 2RT, and 3RT, respectively. Under a compressive load of 1000 N, the displacement decreased with an increasing number of reverse transformation cycles, and the 3RT specimen showed the highest apparent resistance to deformation. The loads required to produce a displacement of 5 mm were 812, 843, 865, and 896 N for the as-built, RT, 2RT, and 3RT rods, respectively; thus, the 3RT rod exhibited an increase of 84 N relative to the as-built rod. Although all specimens were fabricated using the same CAD geometry and adjusted to the same nominal dimensional profile, small specimen-specific geometric deviations could not be independently quantified. Therefore, the observed differences in load–displacement response cannot be attributed exclusively to heat-treatment-induced microstructural changes. Furthermore, increasing the maximum diameter of the middle region of the variable-diameter rod from 7 to 8 mm reduced the displacement under 1000 N from 3.3 to 2.7 mm. In a single-specimen cyclic compression screening test at 0–300 N, 1 Hz, and 1 × 106 cycles, no apparent macroscopic residual deformation or fracture was observed. This result represents runout under the specified loading condition and does not establish a fatigue limit. These results support the feasibility of using PBF-EB to fabricate CCMN rods with prescribed curved and variable-diameter geometries and demonstrate a stable mechanical response under the specific static and cyclic loading conditions examined in this study.

AlloysVol. 5(4)
University of Shanghai for Science and Technology (CN), Tohoku University (JP), New Industry Creation Hatchery Center, Tohoku University (JP), Josai International University (JP)
Good health and well-being
Additive Manufacturing Materials and Processes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon