The Gator low-background spectrometer: readout upgrade and photomultiplier tube radioassay for XLZD
We report on a readout upgrade of Gator, a low-background high-purity germanium spectrometer operated at the Laboratori Nazionali del Gran Sasso for material radioassay. The upgraded data acquisition records digitised waveforms, enabling noise rejection on an event-level basis, and offline energy reconstruction. Measurements before and after the upgrade show an improved energy resolution of 17% at the 2614.5 keV line of 208-Tl, with (2.14 +- 0.22) keV at FWHM, and suppression of electronic noise in the low-energy region. We assess the resulting sensitivity for a specified sample geometry, and present a new background measurement. The background rate in the (100-2700) keV range is (86.6 +- 0.5) counts/(kg day). We also show the radioassay results of sixteen Hamamatsu R17317 3-inch photomultiplier tubes with an improved stem design, which reduces the activities from the primordial 238-U and 232-Th chains by a factor of three, making them suitable candidate photosensors for the XLZD experiment.