The Gator low-background spectrometer: readout upgrade and photomultiplier tube radioassay for XLZD

We report on a readout upgrade of Gator, a low-background high-purity germanium spectrometer operated at the Laboratori Nazionali del Gran Sasso for material radioassay. The upgraded data acquisition records digitised waveforms, enabling noise rejection on an event-level basis, and offline energy reconstruction. Measurements before and after the upgrade show an improved energy resolution of 17% at the 2614.5 keV line of 208-Tl, with (2.14 +- 0.22) keV at FWHM, and suppression of electronic noise in the low-energy region. We assess the resulting sensitivity for a specified sample geometry, and present a new background measurement. The background rate in the (100-2700) keV range is (86.6 +- 0.5) counts/(kg day). We also show the radioassay results of sixteen Hamamatsu R17317 3-inch photomultiplier tubes with an improved stem design, which reduces the activities from the primordial 238-U and 232-Th chains by a factor of three, making them suitable candidate photosensors for the XLZD experiment.

Authors

Publication Details

Journal
arXiv (Cornell University)
Published
2026-11-12
DOI
https://doi.org/10.48550/arxiv.2611.17187
Primary Topic
Radioactive Decay and Measurement Techniques
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

The Gator low-background spectrometer: readout upgrade and photomultiplier tube radioassay for XLZD

Sofia Calgaro, F. Piastra, Florian Joerg, Margherita Noia et al.
arXiv (Cornell University)
Radioactive Decay and Measurement Techniques
preprint

The Gator low-background spectrometer: readout upgrade and photomultiplier tube radioassay for XLZD

Sofia Calgaro, F. Piastra, Florian Joerg, Margherita Noia, Aravind Sreekala, Laura Baudis, Jose Javier Cuenca-Garcia
preprint en

Abstract

We report on a readout upgrade of Gator, a low-background high-purity germanium spectrometer operated at the Laboratori Nazionali del Gran Sasso for material radioassay. The upgraded data acquisition records digitised waveforms, enabling noise rejection on an event-level basis, and offline energy reconstruction. Measurements before and after the upgrade show an improved energy resolution of 17% at the 2614.5 keV line of 208-Tl, with (2.14 +- 0.22) keV at FWHM, and suppression of electronic noise in the low-energy region. We assess the resulting sensitivity for a specified sample geometry, and present a new background measurement. The background rate in the (100-2700) keV range is (86.6 +- 0.5) counts/(kg day). We also show the radioassay results of sixteen Hamamatsu R17317 3-inch photomultiplier tubes with an improved stem design, which reduces the activities from the primordial 238-U and 232-Th chains by a factor of three, making them suitable candidate photosensors for the XLZD experiment.

arXiv (Cornell University)
Radioactive Decay and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon