Effect of ambient temperature on bioimpedance from a wearable sensor patch

Wearable bioimpedance sensors are promising tools for long-term home monitoring, but their measurements may be affected by ambient temperature and anatomical site. This study examined how cold and warm exposure influence local skin impedance measured by a miniaturized patch at four body locations.Twelve healthy volunteers (4 female, 8 male; 18-40 years) wore four Re:Balans ® Insight patches (upper back, upper arm, thigh, lower leg) for approximately 30 h. On two consecutive days, each participant underwent 1 h of exposure to either a cold (4 °C) or warm (37 °C) climate chamber in counterbalanced order, with preand post-exposure baseline periods. Multi-frequency bioimpedance (24 frequencies, 5.8-504 kHz) and skin temperature were recorded continuously and aligned to the 1 h climate chamber period. Two representative frequencies (7.4 kHz, 403.4 kHz) were analyzed for relative impedance change, time trends, and temperature sensitivity ∆Z/∆T . Non-parametric tests were used to compare locations and conditions.Cold exposure decreased skin temperature by about -3.2 to -5.2 • C and increased impedance at all sites, whereas warm exposure increased skin temperature by about +1.6 to +2.4 • C and decreased impedance at all sites. Relative and absolute changes were larger at 7.4 kHz than at 403.4 kHz. Location-dependent differences were small during cold exposure but more pronounced during warm exposure, with a moderate-to-large effect between upper back and lower leg that did not remain statistically significant after multiple comparison correction. All sites showed negative temperature sensitivity; limb locations exhibited higher |∆Z/∆T | and stronger impedance-temperature linearity than the upper back.Local skin impedance measured by wearable patches is sensitive to changes in ambient temperature in a frequency-and site-dependent manner. Extremity sites appear suitable for temperature-or blood-flow-related applications, whereas the upper back may be advantageous where temperature-induced variability should be minimized.

Authors

Institutions

Publication Details

Journal
Medical Engineering & Physics
Published
2026-11-12
DOI
https://doi.org/10.1088/1873-4030/aec16c
Primary Topic
Electrical and Bioimpedance Tomography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Effect of ambient temperature on bioimpedance from a wearable sensor patch

Ole Elvebakk, Jie Hou, Sisay Mebre Abie, Ørjan G. Martinsen et al.
Medical Engineering & Physics
Electrical and Bioimpedance Tomography
article

Effect of ambient temperature on bioimpedance from a wearable sensor patch

Ole Elvebakk, Jie Hou, Sisay Mebre Abie, Ørjan G. Martinsen, Håvard Kalvøy
article en

Abstract

Wearable bioimpedance sensors are promising tools for long-term home monitoring, but their measurements may be affected by ambient temperature and anatomical site. This study examined how cold and warm exposure influence local skin impedance measured by a miniaturized patch at four body locations.Twelve healthy volunteers (4 female, 8 male; 18-40 years) wore four Re:Balans ® Insight patches (upper back, upper arm, thigh, lower leg) for approximately 30 h. On two consecutive days, each participant underwent 1 h of exposure to either a cold (4 °C) or warm (37 °C) climate chamber in counterbalanced order, with preand post-exposure baseline periods. Multi-frequency bioimpedance (24 frequencies, 5.8-504 kHz) and skin temperature were recorded continuously and aligned to the 1 h climate chamber period. Two representative frequencies (7.4 kHz, 403.4 kHz) were analyzed for relative impedance change, time trends, and temperature sensitivity ∆Z/∆T . Non-parametric tests were used to compare locations and conditions.Cold exposure decreased skin temperature by about -3.2 to -5.2 • C and increased impedance at all sites, whereas warm exposure increased skin temperature by about +1.6 to +2.4 • C and decreased impedance at all sites. Relative and absolute changes were larger at 7.4 kHz than at 403.4 kHz. Location-dependent differences were small during cold exposure but more pronounced during warm exposure, with a moderate-to-large effect between upper back and lower leg that did not remain statistically significant after multiple comparison correction. All sites showed negative temperature sensitivity; limb locations exhibited higher |∆Z/∆T | and stronger impedance-temperature linearity than the upper back.Local skin impedance measured by wearable patches is sensitive to changes in ambient temperature in a frequency-and site-dependent manner. Extremity sites appear suitable for temperature-or blood-flow-related applications, whereas the upper back may be advantageous where temperature-induced variability should be minimized.

Medical Engineering & Physics
Oslo University Hospital (NO), University of Oslo (NO)
Electrical and Bioimpedance Tomography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon