Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study
Background: This study aims to establish local diagnostic reference levels (DRLs) for mechanical thrombectomy performed using a low-dose imaging protocol and to assess objective image quality parameters under routine clinical conditions. Methods: This retrospective, single-center observational study included 59 consecutive patients who underwent mechanical thrombectomy for acute ischemic stroke between January 2023 and March 2026, with no exclusions. Radiation metrics, including dose-area product (DAP), reference air kerma (Ka,r), fluoroscopy time (FT), and number of images, were extracted from system-generated dose reports. Local DRLs were defined as the 75th percentile (P75) of the dose distribution. Objective image quality was assessed using quantitative parameters such as background noise, signal-to-noise ratio (SNR), absolute contrast-to-noise ratio (CNR), CNR/DAP, CNR/Ka,r, and figure-of-merit (FOM = CNR2/DAP). Results: In the overall cohort, the median and P75 values were 16.6/26.2 Gy·cm2 for DAP, 206/306 mGy for Ka,r, and 1217/1622 s for FT. The median number of images per procedure was 210, with a P75 of 285. Procedures utilizing combined biplane imaging demonstrated higher DAP and Ka,r values compared to those using single-plane imaging, while FT did not differ significantly. Objective image-quality analysis was feasible in all 59 procedures. The median background noise, SNR, and absolute CNR were 44.24, 34.47, and 26.99, respectively. Median dose-normalized values were 1.32 for CNR/DAP, 0.12 for CNR/Ka,r, and 33.21 for FOM. No procedure required escalation of the imaging protocol due to insufficient angiographic visualization. Conclusions: Mechanical thrombectomy performed under a predefined low-dose protocol yielded low absolute local dose values in this single-center cohort. ROI-based image-quality parameters remained measurable. Because no standard-dose comparator was imaged, the magnitude of dose reduction and the preservation of image quality relative to conventional settings could not be quantified.
Authors
- M. Budzanowski (ORCID: https://orcid.org/0000-0001-9426-7619)
- Mariusz Stanisław Sowa (ORCID: https://orcid.org/0000-0002-8615-6369)
- Joanna Sowa (ORCID: https://orcid.org/0000-0002-2402-5363)
Institutions
- Institute of Nuclear Physics, Polish Academy of Sciences (PL)
- University of Warmia and Mazury in Olsztyn (PL)
Publication Details
- Journal
- Journal of Clinical Medicine
- Published
- 2026-11-12
- DOI
- https://doi.org/10.3390/jcm15228846
- Primary Topic
- Radiation Dose and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00