Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study

Background: This study aims to establish local diagnostic reference levels (DRLs) for mechanical thrombectomy performed using a low-dose imaging protocol and to assess objective image quality parameters under routine clinical conditions. Methods: This retrospective, single-center observational study included 59 consecutive patients who underwent mechanical thrombectomy for acute ischemic stroke between January 2023 and March 2026, with no exclusions. Radiation metrics, including dose-area product (DAP), reference air kerma (Ka,r), fluoroscopy time (FT), and number of images, were extracted from system-generated dose reports. Local DRLs were defined as the 75th percentile (P75) of the dose distribution. Objective image quality was assessed using quantitative parameters such as background noise, signal-to-noise ratio (SNR), absolute contrast-to-noise ratio (CNR), CNR/DAP, CNR/Ka,r, and figure-of-merit (FOM = CNR2/DAP). Results: In the overall cohort, the median and P75 values were 16.6/26.2 Gy·cm2 for DAP, 206/306 mGy for Ka,r, and 1217/1622 s for FT. The median number of images per procedure was 210, with a P75 of 285. Procedures utilizing combined biplane imaging demonstrated higher DAP and Ka,r values compared to those using single-plane imaging, while FT did not differ significantly. Objective image-quality analysis was feasible in all 59 procedures. The median background noise, SNR, and absolute CNR were 44.24, 34.47, and 26.99, respectively. Median dose-normalized values were 1.32 for CNR/DAP, 0.12 for CNR/Ka,r, and 33.21 for FOM. No procedure required escalation of the imaging protocol due to insufficient angiographic visualization. Conclusions: Mechanical thrombectomy performed under a predefined low-dose protocol yielded low absolute local dose values in this single-center cohort. ROI-based image-quality parameters remained measurable. Because no standard-dose comparator was imaged, the magnitude of dose reduction and the preservation of image quality relative to conventional settings could not be quantified.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2026-11-12
DOI
https://doi.org/10.3390/jcm15228846
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study

M. Budzanowski, Mariusz Stanisław Sowa, Joanna Sowa
Journal of Clinical Medicine
Radiation Dose and Imaging
article

Local Diagnostic Reference Levels and Objective Image Quality in Mechanical Thrombectomy Performed with a Low-Dose Protocol: A Single-Center Observational Study

M. Budzanowski, Mariusz Stanisław Sowa, Joanna Sowa
article en

Abstract

Background: This study aims to establish local diagnostic reference levels (DRLs) for mechanical thrombectomy performed using a low-dose imaging protocol and to assess objective image quality parameters under routine clinical conditions. Methods: This retrospective, single-center observational study included 59 consecutive patients who underwent mechanical thrombectomy for acute ischemic stroke between January 2023 and March 2026, with no exclusions. Radiation metrics, including dose-area product (DAP), reference air kerma (Ka,r), fluoroscopy time (FT), and number of images, were extracted from system-generated dose reports. Local DRLs were defined as the 75th percentile (P75) of the dose distribution. Objective image quality was assessed using quantitative parameters such as background noise, signal-to-noise ratio (SNR), absolute contrast-to-noise ratio (CNR), CNR/DAP, CNR/Ka,r, and figure-of-merit (FOM = CNR2/DAP). Results: In the overall cohort, the median and P75 values were 16.6/26.2 Gy·cm2 for DAP, 206/306 mGy for Ka,r, and 1217/1622 s for FT. The median number of images per procedure was 210, with a P75 of 285. Procedures utilizing combined biplane imaging demonstrated higher DAP and Ka,r values compared to those using single-plane imaging, while FT did not differ significantly. Objective image-quality analysis was feasible in all 59 procedures. The median background noise, SNR, and absolute CNR were 44.24, 34.47, and 26.99, respectively. Median dose-normalized values were 1.32 for CNR/DAP, 0.12 for CNR/Ka,r, and 33.21 for FOM. No procedure required escalation of the imaging protocol due to insufficient angiographic visualization. Conclusions: Mechanical thrombectomy performed under a predefined low-dose protocol yielded low absolute local dose values in this single-center cohort. ROI-based image-quality parameters remained measurable. Because no standard-dose comparator was imaged, the magnitude of dose reduction and the preservation of image quality relative to conventional settings could not be quantified.

Journal of Clinical MedicineVol. 15(22)
Institute of Nuclear Physics, Polish Academy of Sciences (PL), University of Warmia and Mazury in Olsztyn (PL)
Good health and well-being
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon