Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison

Background: Maximal inspiratory (MIP) and expiratory (MEP) pressures are widely used indices of respiratory muscle strength, but clinical manometers are costly. We evaluated the bench agreement of two units of a low-cost digital manometer with a clinical respiratory pressure comparator. Methods: Two BioPimo 101® units and a Vitalograph Pneumotrac 6800® were connected simultaneously to a mechanical insufflation–exsufflation device coupled to a test lung. Pressures of 5–70 cmH2O were applied in 5-cmH2O steps under two compliance conditions, in triplicate (56 conditions; 336 paired differences analyzed as repeated measurements). Agreement was assessed with Bland–Altman analysis, condition-clustered bootstrap confidence intervals, regression-based limits of agreement and ICC (2,1). Results: For MIP, mean differences were −0.03 to 0.21 cmH2O (limits of agreement −0.42 to 0.72 cmH2O). For MEP, both units showed higher readings than the comparator (mean differences 0.52–0.72 cmH2O), and differences increased with pressure; at 70 cmH2O under low compliance, the predicted upper limit of agreement reached 2.44–2.64 cmH2O. ICCs exceeded 0.999. Conclusions: Both units agreed closely with the comparator under bench conditions, particularly for MIP, whereas MEP showed positive, pressure-dependent differences. Without a traceable pressure standard and with only two units tested, these findings do not establish absolute accuracy or clinical interchangeability.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-11-12
DOI
https://doi.org/10.3390/diagnostics16223745
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison

Matías Otto‐Yáñez, Homero Puppo G., Marisol Barros-Poblete, Luis Vasconcello‐Castillo et al.
Diagnostics
Respiratory Support and Mechanisms
article

Bench Agreement of Two Units of a Low-Cost Digital Manometers with a Clinical Respiratory Pressure Comparator: A Laboratory-Based Comparison

Matías Otto‐Yáñez, Homero Puppo G., Marisol Barros-Poblete, Luis Vasconcello‐Castillo, Rodrigo Torres‐Castro, Roberto Vera‐Uribe, Giovanna Binetti, Claudia Andrea de los Angeles Saldaña Villegas
article en

Abstract

Background: Maximal inspiratory (MIP) and expiratory (MEP) pressures are widely used indices of respiratory muscle strength, but clinical manometers are costly. We evaluated the bench agreement of two units of a low-cost digital manometer with a clinical respiratory pressure comparator. Methods: Two BioPimo 101® units and a Vitalograph Pneumotrac 6800® were connected simultaneously to a mechanical insufflation–exsufflation device coupled to a test lung. Pressures of 5–70 cmH2O were applied in 5-cmH2O steps under two compliance conditions, in triplicate (56 conditions; 336 paired differences analyzed as repeated measurements). Agreement was assessed with Bland–Altman analysis, condition-clustered bootstrap confidence intervals, regression-based limits of agreement and ICC (2,1). Results: For MIP, mean differences were −0.03 to 0.21 cmH2O (limits of agreement −0.42 to 0.72 cmH2O). For MEP, both units showed higher readings than the comparator (mean differences 0.52–0.72 cmH2O), and differences increased with pressure; at 70 cmH2O under low compliance, the predicted upper limit of agreement reached 2.44–2.64 cmH2O. ICCs exceeded 0.999. Conclusions: Both units agreed closely with the comparator under bench conditions, particularly for MIP, whereas MEP showed positive, pressure-dependent differences. Without a traceable pressure standard and with only two units tested, these findings do not establish absolute accuracy or clinical interchangeability.

DiagnosticsVol. 16(22)
Universidad Autónoma de Chile (CL), Universidad Santo Tomás (CL), Universidad Norbert Wiener (PE), Instituto Ecuatoriano de Enfermedades Digestivas (EC), Private University of the North (PE), University of Chile (CL)
Good health and well-being
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon