Improved Reproducibility of FT-ICR Mass Spectra for Organic Matter by Minimizing Electronic Noise Peaks and Low-abundance Peaks with Few Occurrences

21 Peak reproducibility is a major factor that affects the reliability of Fourier transform 22 ion cyclotron resonance mass spectrometry (FT-ICR MS) spectra of dissolved organic 23 matter (DOM), which was addressed by a newly proposed workflow, namely parallel 24 samples representation by similar samples (PSRSS). The performance of the workflow was 25 validated by parallel measurements and experimental replicates. Inspection of isotopic and 26 homologous peaks in the PSRSS workflow enabled elimination of electronic noise peaks 27 and low-abundance peaks with few occurrences, significantly decreased the Bray-Curtis 28 dissimilarity, and improved the molecular formula assignment ratio with only minor effects 29 on the assigned formulae. The Bray-Curtis dissimilarity threshold (55% to 60% or quartile) 30 was proposed to cluster FT-ICR MS spectra for DOM samples with similar molecular 31 composition. The noise peak removal performance was further improved for the similar 32 molecular composition samples by excluding outlier spectra and less frequently occurring 33 peaks via the PSRSS workflow. Moreover, the reliability of DOM molecular 34 characterization was significantly improved by the PSRSS workflow. Overall, our findings 35 demonstrated the feasibility of the PSRSS workflow to serve as a reliable, labor-saving 36 method to remove electronic noise peaks and cluster spectra, yielding reliable spectral 37 information for multiple independent FT-ICR MS spectra. 38

Authors

Institutions

Publication Details

Journal
ChemRxiv
Published
2026-11-12
DOI
https://doi.org/10.26434/chemrxiv.15012069/v2
Primary Topic
Mass Spectrometry Techniques and Applications
Type
preprint

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Improved Reproducibility of FT-ICR Mass Spectra for Organic Matter by Minimizing Electronic Noise Peaks and Low-abundance Peaks with Few Occurrences

Qing‐Long Fu, Lee M. Blaney, Manabu Fujii, Pingqing Fu et al.
ChemRxiv
Mass Spectrometry Techniques and Applications
preprint

Improved Reproducibility of FT-ICR Mass Spectra for Organic Matter by Minimizing Electronic Noise Peaks and Low-abundance Peaks with Few Occurrences

Qing‐Long Fu, Lee M. Blaney, Manabu Fujii, Pingqing Fu, Xinyi Chen, Qingqing Huang, Yanhui Cheng, Yanxin Wang, Dongmei Zhou, Huiyu Dong, Rui Ma
preprint en

Abstract

21 Peak reproducibility is a major factor that affects the reliability of Fourier transform 22 ion cyclotron resonance mass spectrometry (FT-ICR MS) spectra of dissolved organic 23 matter (DOM), which was addressed by a newly proposed workflow, namely parallel 24 samples representation by similar samples (PSRSS). The performance of the workflow was 25 validated by parallel measurements and experimental replicates. Inspection of isotopic and 26 homologous peaks in the PSRSS workflow enabled elimination of electronic noise peaks 27 and low-abundance peaks with few occurrences, significantly decreased the Bray-Curtis 28 dissimilarity, and improved the molecular formula assignment ratio with only minor effects 29 on the assigned formulae. The Bray-Curtis dissimilarity threshold (55% to 60% or quartile) 30 was proposed to cluster FT-ICR MS spectra for DOM samples with similar molecular 31 composition. The noise peak removal performance was further improved for the similar 32 molecular composition samples by excluding outlier spectra and less frequently occurring 33 peaks via the PSRSS workflow. Moreover, the reliability of DOM molecular 34 characterization was significantly improved by the PSRSS workflow. Overall, our findings 35 demonstrated the feasibility of the PSRSS workflow to serve as a reliable, labor-saving 36 method to remove electronic noise peaks and cluster spectra, yielding reliable spectral 37 information for multiple independent FT-ICR MS spectra. 38

ChemRxiv
Tianjin University (CN), China University of Geosciences (CN), State Key Laboratory of Pollution Control and Resource Reuse (CN), Research Center for Eco-Environmental Sciences (CN), Institute of Science Tokyo (JP), University of Maryland, College Park (US), Nanjing University (CN)
National Natural Science Foundation of China, Ministry of Education of the People's Republic of China
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon