Calculating peptide concentration from UV absorbance at 205 nm or 280 nm v1

This protocol converts a measured UV absorbance at 205 nm or 280 nm into the molar and mass concentration of a peptide in solution, using the Beer-Lambert law with sequence-derived molar absorption coefficients and average masses from the open peptide reference data published by British Peptide Labs (CC BY 4.0). It gives the arithmetic steps and a worked example. For research use only. Not for human or veterinary use.

Authors

Publication Details

Published
2026-11-12
DOI
https://doi.org/10.17504/protocols.io.epv567534g1b/v1
Primary Topic
Chemical Synthesis and Analysis
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Calculating peptide concentration from UV absorbance at 205 nm or 280 nm v1

Sullivan Butt
Chemical Synthesis and Analysis
preprint

Calculating peptide concentration from UV absorbance at 205 nm or 280 nm v1

Sullivan Butt
preprint en

Abstract

This protocol converts a measured UV absorbance at 205 nm or 280 nm into the molar and mass concentration of a peptide in solution, using the Beer-Lambert law with sequence-derived molar absorption coefficients and average masses from the open peptide reference data published by British Peptide Labs (CC BY 4.0). It gives the arithmetic steps and a worked example. For research use only. Not for human or veterinary use.

Chemical Synthesis and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon