REPEATED PRESSURE-RELATED CHROMATOGRAPHIC FAILURE DURING SALIVARY CORTISOL ANALYSIS BY LC-HRMS/MS: A LABORATORY CASE REPORT AND HYPOTHESIS-DRIVEN ROOT CAUSE ASSESSMENT

Liquid chromatography tandem mass spectrometry (LC-MS/MS) is the reference standard method for quantitatively measuring salivary cortisol, largely because of its high analytical specificity and sensitivity. However, saliva is a complex biological matrix that may compromise chromatographic robustness during routine testing. Here, we report a laboratory case of system back-pressure exceedance, chromatographic deterioration, and analytical column collapse during routine LC-HRMS/MS analysis of salivary cortisol using a commercially available sample preparation and analysis kit. Two columns used in assay analysis failed consecutively. Both columns were subjected to the same working conditions. Factors such as how the samples were injected, chromatograms, and the behaviour of the system were analyzed. Possible causes of failures were also evaluated based on the pressure and behaviour of the columns. The team also evaluated possible causes based on column pressure and behaviour. The review indicated that the matrix caused the failure. Poorly removed salivary mucins and other particles clogged the columns. Flow restriction caused retention time to vary. The columns became worn out because the phase failed to remain permanently fixed to the column. Failures occurred in two columns and suggested a limitation of the method regarding matrix interaction with LC-MS/MS. This occurred mainly when high-viscosity saliva samples were analyzed using UHPLC columns. Restrictions were noted particularly at the column inlet. It was recommended that columns be well protected and that preventive maintenance be performed to reduce column failures and loss of phase integrity. This is especially true in laboratories that analyze salivary cortisol using LC-MS/MS. The report presents a case study illustrating the types of column failures caused by matrix limitations in LC-MS/MS analysis.

Authors

Publication Details

Journal
Bulletin of Biological and Allied Sciences Research
Published
2026-11-12
DOI
https://doi.org/10.64013/bbasr.v2026i1.141
Primary Topic
Analytical Chemistry and Chromatography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

REPEATED PRESSURE-RELATED CHROMATOGRAPHIC FAILURE DURING SALIVARY CORTISOL ANALYSIS BY LC-HRMS/MS: A LABORATORY CASE REPORT AND HYPOTHESIS-DRIVEN ROOT CAUSE ASSESSMENT

SR UMRAO, SHS MOHAMMED, F ASHRAF, MR QURESHI et al.
Bulletin of Biological and Allied Sciences Research
Analytical Chemistry and Chromatography
article

REPEATED PRESSURE-RELATED CHROMATOGRAPHIC FAILURE DURING SALIVARY CORTISOL ANALYSIS BY LC-HRMS/MS: A LABORATORY CASE REPORT AND HYPOTHESIS-DRIVEN ROOT CAUSE ASSESSMENT

SR UMRAO, SHS MOHAMMED, F ASHRAF, MR QURESHI, FU AMIN, MS THARAMMAL, MM KHAN, S ELANGOVAN, MM ALAM
article en

Abstract

Liquid chromatography tandem mass spectrometry (LC-MS/MS) is the reference standard method for quantitatively measuring salivary cortisol, largely because of its high analytical specificity and sensitivity. However, saliva is a complex biological matrix that may compromise chromatographic robustness during routine testing. Here, we report a laboratory case of system back-pressure exceedance, chromatographic deterioration, and analytical column collapse during routine LC-HRMS/MS analysis of salivary cortisol using a commercially available sample preparation and analysis kit. Two columns used in assay analysis failed consecutively. Both columns were subjected to the same working conditions. Factors such as how the samples were injected, chromatograms, and the behaviour of the system were analyzed. Possible causes of failures were also evaluated based on the pressure and behaviour of the columns. The team also evaluated possible causes based on column pressure and behaviour. The review indicated that the matrix caused the failure. Poorly removed salivary mucins and other particles clogged the columns. Flow restriction caused retention time to vary. The columns became worn out because the phase failed to remain permanently fixed to the column. Failures occurred in two columns and suggested a limitation of the method regarding matrix interaction with LC-MS/MS. This occurred mainly when high-viscosity saliva samples were analyzed using UHPLC columns. Restrictions were noted particularly at the column inlet. It was recommended that columns be well protected and that preventive maintenance be performed to reduce column failures and loss of phase integrity. This is especially true in laboratories that analyze salivary cortisol using LC-MS/MS. The report presents a case study illustrating the types of column failures caused by matrix limitations in LC-MS/MS analysis.

Bulletin of Biological and Allied Sciences ResearchVol. 2026(1)
Good health and well-being
Analytical Chemistry and Chromatography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon