Clinical Validation of Virtual Reality–Based Oculography for Quantitative Ocular Alignment Assessment in Strabismus

Abstract Purpose To evaluate the clinical validity of a virtual reality (VR)-based eye-tracking system for quantitative ocular alignment assessment by comparing device-derived measurements with conventional clinical measurements in strabismus. Methods We retrospectively reviewed medical records of consecutive patients undergoing strabismus evaluation. We assessed ocular alignment using conventional prism cover testing and a VR-based oculography system (neos™, machineMD, Bern, Switzerland). We compared total deviation measured by alternate prism cover testing (APCT) and manifest deviation measured by simultaneous prism cover testing (SPCT) with device-derived measurements. We assessed associations using Spearman correlation and agreement using Bland–Altman analysis. Subgroup analyses were performed by horizontal deviation magnitude (<15 vs. ≥15 prism diopters [PD]). Results Forty-six patients (mean age, 37.61 ± 20.41 years; 25 [54.3%] female) were included. Device-derived measurements correlated significantly with conventional measurements for horizontal total deviation (Spearman ρ = 0.780, P < 0.001), vertical total deviation (ρ = 0.728, P < 0.001), horizontal tropia (ρ = 0.783, P < 0.001), and vertical tropia (ρ = 0.424, P = 0.003). Bland–Altman analysis showed closer agreement for horizontal than vertical measurements. Measurement differences did not differ significantly between the <15 PD and ≥15 PD groups, although greater variability was observed in the ≥15 PD group. Conclusions VR-based oculography demonstrated significant associations with conventional measurements, with better agreement for horizontal than vertical deviations. Greater variability with larger horizontal deviations suggests cautious interpretation in large-angle strabismus.

Authors

Institutions

Publication Details

Journal
Research Square
Published
2026-11-11
DOI
https://doi.org/10.21203/rs.3.rs-10922714/v1
Primary Topic
Ophthalmology and Eye Disorders
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Clinical Validation of Virtual Reality–Based Oculography for Quantitative Ocular Alignment Assessment in Strabismus

Donny Won Suh, Kimia Rezaei, Donghun Lee, Mia Jia Wang
Research Square
Ophthalmology and Eye Disorders
preprint

Clinical Validation of Virtual Reality–Based Oculography for Quantitative Ocular Alignment Assessment in Strabismus

Donny Won Suh, Kimia Rezaei, Donghun Lee, Mia Jia Wang
preprint en

Abstract

Abstract Purpose To evaluate the clinical validity of a virtual reality (VR)-based eye-tracking system for quantitative ocular alignment assessment by comparing device-derived measurements with conventional clinical measurements in strabismus. Methods We retrospectively reviewed medical records of consecutive patients undergoing strabismus evaluation. We assessed ocular alignment using conventional prism cover testing and a VR-based oculography system (neos™, machineMD, Bern, Switzerland). We compared total deviation measured by alternate prism cover testing (APCT) and manifest deviation measured by simultaneous prism cover testing (SPCT) with device-derived measurements. We assessed associations using Spearman correlation and agreement using Bland–Altman analysis. Subgroup analyses were performed by horizontal deviation magnitude (<15 vs. ≥15 prism diopters [PD]). Results Forty-six patients (mean age, 37.61 ± 20.41 years; 25 [54.3%] female) were included. Device-derived measurements correlated significantly with conventional measurements for horizontal total deviation (Spearman ρ = 0.780, P < 0.001), vertical total deviation (ρ = 0.728, P < 0.001), horizontal tropia (ρ = 0.783, P < 0.001), and vertical tropia (ρ = 0.424, P = 0.003). Bland–Altman analysis showed closer agreement for horizontal than vertical measurements. Measurement differences did not differ significantly between the <15 PD and ≥15 PD groups, although greater variability was observed in the ≥15 PD group. Conclusions VR-based oculography demonstrated significant associations with conventional measurements, with better agreement for horizontal than vertical deviations. Greater variability with larger horizontal deviations suggests cautious interpretation in large-angle strabismus.

Research Square
University of California, Irvine (US), Daegu Catholic University (KR)
Ophthalmology and Eye Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon