A Source-Traced PET/CT Scanner Performance Database (2004–2024): The Sphere-Size Dependence of Time-of-Flight Contrast Recovery

Background: Small-lesion quantification is a key task in diagnostic PET/CT. However, published performance data span two decades and multiple NEMA NU 2 revisions, complicating comparisons. We developed a source-traced database to examine how reconstruction-associated improvements in hot-sphere contrast recovery vary with sphere diameter, providing reference data for phantom-based protocol optimisation. Methods: Published NEMA and phantom measurements were source-verified, linked to scanner geometry, and annotated with acquisition and reconstruction conditions. Contrast recovery at a 4:1 sphere-to-background activity ratio (151 measurements; 13 scanners) was analysed using mixed-effects meta-regression with a scanner random intercept. Results: The database included 28 scanner models (2004–2024), 610 measurements and 43 publications. Descriptively, sensitivity was 3.8–10.9 cps/kBq for lutetium-based analogue scanners, 12.1–47.6 for conventional-geometry digital scanners and 176–192 for total-body scanners. Time of flight (TOF) was associated with a 7.9-percentage-point-higher contrast recovery (95% CI 4.8 to 11.1), but the association varied with sphere diameter (p < 0.001) and was significant only at 10–17 mm (11.4, 9.9 and 7.0 points); estimates at 28 and 37 mm rested on four to five scanners and were imprecise. The median between-laboratory difference under matched conditions was 5.7 points. Point-spread-function and regularised-reconstruction effects were not robust. Conclusions: In phantom measurements, TOF-associated gains were concentrated at 10–17 mm. These data provide reference values for phantom-based protocol optimisation, harmonisation and acceptance testing; they do not measure lesion detectability or diagnostic accuracy, which require patient-based validation.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-11-11
DOI
https://doi.org/10.3390/diagnostics16223725
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A Source-Traced PET/CT Scanner Performance Database (2004–2024): The Sphere-Size Dependence of Time-of-Flight Contrast Recovery

Rajeh Assiri, Mohammad Sayed, Ibrahim Hadadi, Hussein Al‐shaari et al.
Diagnostics
Medical Imaging Techniques and Applications
article

A Source-Traced PET/CT Scanner Performance Database (2004–2024): The Sphere-Size Dependence of Time-of-Flight Contrast Recovery

Rajeh Assiri, Mohammad Sayed, Ibrahim Hadadi, Hussein Al‐shaari, Mansour M. Alqahtani, Mohammed Asiri, Abdulaziz Yahya M. Hussain
article en

Abstract

Background: Small-lesion quantification is a key task in diagnostic PET/CT. However, published performance data span two decades and multiple NEMA NU 2 revisions, complicating comparisons. We developed a source-traced database to examine how reconstruction-associated improvements in hot-sphere contrast recovery vary with sphere diameter, providing reference data for phantom-based protocol optimisation. Methods: Published NEMA and phantom measurements were source-verified, linked to scanner geometry, and annotated with acquisition and reconstruction conditions. Contrast recovery at a 4:1 sphere-to-background activity ratio (151 measurements; 13 scanners) was analysed using mixed-effects meta-regression with a scanner random intercept. Results: The database included 28 scanner models (2004–2024), 610 measurements and 43 publications. Descriptively, sensitivity was 3.8–10.9 cps/kBq for lutetium-based analogue scanners, 12.1–47.6 for conventional-geometry digital scanners and 176–192 for total-body scanners. Time of flight (TOF) was associated with a 7.9-percentage-point-higher contrast recovery (95% CI 4.8 to 11.1), but the association varied with sphere diameter (p < 0.001) and was significant only at 10–17 mm (11.4, 9.9 and 7.0 points); estimates at 28 and 37 mm rested on four to five scanners and were imprecise. The median between-laboratory difference under matched conditions was 5.7 points. Point-spread-function and regularised-reconstruction effects were not robust. Conclusions: In phantom measurements, TOF-associated gains were concentrated at 10–17 mm. These data provide reference values for phantom-based protocol optimisation, harmonisation and acceptance testing; they do not measure lesion detectability or diagnostic accuracy, which require patient-based validation.

DiagnosticsVol. 16(22)
Taif University (SA), Najran University (SA), King Khalid University (SA)
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon