Comparison of Cone-Beam CT, Photon-Counting Detector CT and Micro-CT for Imaging Bone Graft Biomaterials and Three-Dimensional Printed Calcium Phosphate Scaffolds: An Ex Vivo Study

Background/Objectives: Cone-beam computed tomography (CBCT) is widely used in dentomaxillofacial imaging for evaluating bone grafts and implant-related procedures; however, image quality is often compromised by metal-induced artefacts. This study compared CBCT, photon-counting detector CT (PCD-CT), and micro-computed tomography (µCT) for imaging clinically relevant bone graft materials and three-dimensional printed calcium phosphate scaffolds, focusing on biomaterial visualisation and titanium artefact reduction. Methods: Four bovine mandibles were prepared with standardized cylindrical defects containing eight bone substitute materials and 3D-printed calcium phosphate scaffolds, with or without surrounding titanium cages. Specimens were imaged using CBCT and PCD-CT with low-dose and maximum-dose ultra-high-resolution acquisition protocols, and µCT. Two readers independently assessed image interpretability, biomaterial visualisation, and metal artefact severity using a five-point scoring system. Inter-reader agreement was evaluated using weighted Cohen’s kappa. Results: µCT provided the highest spatial resolution and served as the ex vivo reference standard. Under the acquisition and reconstruction protocols evaluated, PCD-CT provided favorable visualization of scaffold architecture and reduced the visual impact of titanium-induced artefacts compared with CBCT. High-energy virtual monoenergetic reconstructions (≥130 keV) further reduced beam-hardening and streak artefacts, although the optimal reconstruction energy may depend on the biomaterial and diagnostic task. Without titanium cages, PCD-CT and µCT achieved higher image interpretability scores than CBCT. With titanium cages, CBCT frequently became non-diagnostic, whereas PCD-CT maintained good visualisation of most biomaterials. Inter-reader agreement ranged from substantial to almost perfect. Conclusions: In this exploratory ex vivo study, the evaluated PCD-CT protocols provided favorable qualitative visualization of several bone graft materials and three-dimensional printed calcium phosphate scaffolds compared with the evaluated CBCT protocol, particularly in the presence of titanium cages. However, the protocols were not dose-matched, and differences in acquisition and reconstruction parameters prevent attributing the observed findings exclusively to detector technology. The study does not establish PCD-CT superiority at equivalent radiation doses or demonstrate clinical performance in longitudinal bone-regeneration monitoring. Further dose-matched studies incorporating objective image-quality measurements and clinical validation are required.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-11-11
DOI
https://doi.org/10.3390/diagnostics16223728
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Comparison of Cone-Beam CT, Photon-Counting Detector CT and Micro-CT for Imaging Bone Graft Biomaterials and Three-Dimensional Printed Calcium Phosphate Scaffolds: An Ex Vivo Study

Richard Frank Richter, Tristan T. Demmert, Adib Al‐Haj Husain, Bernd Stadlinger et al.
Diagnostics
Advanced X-ray and CT Imaging
article

Comparison of Cone-Beam CT, Photon-Counting Detector CT and Micro-CT for Imaging Bone Graft Biomaterials and Three-Dimensional Printed Calcium Phosphate Scaffolds: An Ex Vivo Study

Richard Frank Richter, Tristan T. Demmert, Adib Al‐Haj Husain, Bernd Stadlinger, Harald Essig, Julien Guerrero, Maximilian E.H. Wagner, Hatem Alkadhi, Victor Mergen, Franz E. Weber, Thomas G. Flohr, Paula Korn, Ana Pérez Domínguez
article en

Abstract

Background/Objectives: Cone-beam computed tomography (CBCT) is widely used in dentomaxillofacial imaging for evaluating bone grafts and implant-related procedures; however, image quality is often compromised by metal-induced artefacts. This study compared CBCT, photon-counting detector CT (PCD-CT), and micro-computed tomography (µCT) for imaging clinically relevant bone graft materials and three-dimensional printed calcium phosphate scaffolds, focusing on biomaterial visualisation and titanium artefact reduction. Methods: Four bovine mandibles were prepared with standardized cylindrical defects containing eight bone substitute materials and 3D-printed calcium phosphate scaffolds, with or without surrounding titanium cages. Specimens were imaged using CBCT and PCD-CT with low-dose and maximum-dose ultra-high-resolution acquisition protocols, and µCT. Two readers independently assessed image interpretability, biomaterial visualisation, and metal artefact severity using a five-point scoring system. Inter-reader agreement was evaluated using weighted Cohen’s kappa. Results: µCT provided the highest spatial resolution and served as the ex vivo reference standard. Under the acquisition and reconstruction protocols evaluated, PCD-CT provided favorable visualization of scaffold architecture and reduced the visual impact of titanium-induced artefacts compared with CBCT. High-energy virtual monoenergetic reconstructions (≥130 keV) further reduced beam-hardening and streak artefacts, although the optimal reconstruction energy may depend on the biomaterial and diagnostic task. Without titanium cages, PCD-CT and µCT achieved higher image interpretability scores than CBCT. With titanium cages, CBCT frequently became non-diagnostic, whereas PCD-CT maintained good visualisation of most biomaterials. Inter-reader agreement ranged from substantial to almost perfect. Conclusions: In this exploratory ex vivo study, the evaluated PCD-CT protocols provided favorable qualitative visualization of several bone graft materials and three-dimensional printed calcium phosphate scaffolds compared with the evaluated CBCT protocol, particularly in the presence of titanium cages. However, the protocols were not dose-matched, and differences in acquisition and reconstruction parameters prevent attributing the observed findings exclusively to detector technology. The study does not establish PCD-CT superiority at equivalent radiation doses or demonstrate clinical performance in longitudinal bone-regeneration monitoring. Further dose-matched studies incorporating objective image-quality measurements and clinical validation are required.

DiagnosticsVol. 16(22)
University of Zurich (CH), Maastricht University Medical Centre (NL), University Hospital of Zurich (CH), University Hospital Carl Gustav Carus (DE), Technische Universität Dresden (DE)
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon