A Helium Recovery Condenser for a Low-Temperature Laboratory: Liquefaction Rate, Cold-Head Load and Recovery Fraction

This paper presents a computed design for a helium recovery condenser intended for cryogen recovery systems, with its performance, resource budgets and margins derived from stated assumptions rather than asserted. The headline results are: liquefaction rate set by the limiting stage of 20.3 litres per day; total heat to be removed per litre of liquid produced, all stages of 51.8 kJ per litre; fraction of the laboratory boil-off returned as liquid of 0.813 dimensionless; liquid recovered per day against the stated boil-off of 20.3 litres per day; electrical energy per litre recovered at full rate of 8.85 kWh per litre. Each figure was produced by the deposited computation from the inputs listed in the Methods, and the sensitivity of the conclusions to those inputs is examined in the Discussion. The design is a paper study: it is labelled, on its face and in this abstract, SIMULATED, NOT MEASURED, and no quantity reported here is a measurement of built hardware.

Authors

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2026-11-12
DOI
https://doi.org/10.5281/zenodo.24147811
Citations
1
Primary Topic
Spacecraft and Cryogenic Technologies
Type
article
Field-Weighted Citation Impact
1.98
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A Helium Recovery Condenser for a Low-Temperature Laboratory: Liquefaction Rate, Cold-Head Load and Recovery Fraction

William Alexander Aldred
1 citations
Zenodo (CERN European Organization for Nuclear Research)
Spacecraft and Cryogenic Technologies
1.98
article

A Helium Recovery Condenser for a Low-Temperature Laboratory: Liquefaction Rate, Cold-Head Load and Recovery Fraction

William Alexander Aldred
article en
1 citations

Abstract

This paper presents a computed design for a helium recovery condenser intended for cryogen recovery systems, with its performance, resource budgets and margins derived from stated assumptions rather than asserted. The headline results are: liquefaction rate set by the limiting stage of 20.3 litres per day; total heat to be removed per litre of liquid produced, all stages of 51.8 kJ per litre; fraction of the laboratory boil-off returned as liquid of 0.813 dimensionless; liquid recovered per day against the stated boil-off of 20.3 litres per day; electrical energy per litre recovered at full rate of 8.85 kWh per litre. Each figure was produced by the deposited computation from the inputs listed in the Methods, and the sensitivity of the conclusions to those inputs is examined in the Discussion. The design is a paper study: it is labelled, on its face and in this abstract, SIMULATED, NOT MEASURED, and no quantity reported here is a measurement of built hardware.

Zenodo (CERN European Organization for Nuclear Research)
Openalex Percentile: Top 21%
Spacecraft and Cryogenic Technologies
1.98
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
A Helium Recovery Condenser for a Low-Temperature Laboratory: Liquefaction Rate, Cold-Head Load and Recovery Fraction — William Alexander Aldred · Zenodo (CERN European Organization for Nuclear Research) (2026) | TGRS Research Map | TGRS