Challenges in measuring serum calprotectin in autoinflammation: a comparative study of different immunoassays

BACKGROUND: Serum calprotectin (S100A8/A9 complex) is increasingly used in the clinical evaluation of autoinflammatory diseases (AID). However, systematic comparisons of commercially available assays remain limited. This study aimed to assess comparability, and clinical applicability of commercial calprotectin assays to support standardized biomarker-guided diagnostics. METHODS: Serum samples from 40 individuals-systemic juvenile idiopathic arthritis (sJIA, n = 10), adult-onset Still's disease (AOSD, n = 10), familial Mediterranean fever (FMF, n = 10), PAPA syndrome (n = 2), infections (n = 5), leukemia (n = 3), and healthy pediatric controls (n = 10)-were analyzed. Calprotectin concentrations were measured using four ELISAs (BÜHLMANN, R&D Systems, BioLegend, BMA), two PETIAs (sCAL turbo, GCAL), and one CLIA (QuantaFlash) and compared with a BÜHLMANN ELISA reference. ELISA assays were measured with a secondary 10-fold dilution, while PETIA and CLIA assays were performed on automated platforms with reflex dilution. Spiking experiments were conducted to evaluate recovery. RESULTS: All assays except BMA (r = 0.52) showed strong correlation with the reference (r ≥ 0.96). Median values in healthy controls varied markedly (0.29-3.22 µg/mL), with inter-assay differences exceeding one log scale. Several assays underestimated very high concentrations (> 50 µg/mL), likely due to Hook effects, and upper detection limits differed substantially. Spike recovery ranged from 79 to 268%, with the most consistent results observed for sCAL turbo. CONCLUSIONS: Commercial assays reliably quantify serum calprotectin in the normal to moderately elevated range but are not interchangeable. Variability in detection limits and susceptibility to Hook effects may lead to underestimation in severe inflammation. Assay-specific reference ranges and additional dilution steps are recommended, particularly when extreme elevations are suspected.

Authors

Institutions

Publication Details

Journal
Molecular and Cellular Pediatrics
Published
2026-11-10
DOI
https://doi.org/10.1186/s40348-026-00267-8
Primary Topic
S100 Proteins and Annexins
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Challenges in measuring serum calprotectin in autoinflammation: a comparative study of different immunoassays

Andrea Karolin, Michael Peter Horn, Susanne Schleifenbaum, Carolin Park et al.
Molecular and Cellular Pediatrics
S100 Proteins and Annexins
article

Challenges in measuring serum calprotectin in autoinflammation: a comparative study of different immunoassays

Andrea Karolin, Michael Peter Horn, Susanne Schleifenbaum, Carolin Park, Dirk Foell, Manfred Fobker, Helmut Wittkowski, Claas H. Hinze, Melanie Saers, Bernhard Schlüter, Sabrina Fühner
article en

Abstract

BACKGROUND: Serum calprotectin (S100A8/A9 complex) is increasingly used in the clinical evaluation of autoinflammatory diseases (AID). However, systematic comparisons of commercially available assays remain limited. This study aimed to assess comparability, and clinical applicability of commercial calprotectin assays to support standardized biomarker-guided diagnostics. METHODS: Serum samples from 40 individuals-systemic juvenile idiopathic arthritis (sJIA, n = 10), adult-onset Still's disease (AOSD, n = 10), familial Mediterranean fever (FMF, n = 10), PAPA syndrome (n = 2), infections (n = 5), leukemia (n = 3), and healthy pediatric controls (n = 10)-were analyzed. Calprotectin concentrations were measured using four ELISAs (BÜHLMANN, R&D Systems, BioLegend, BMA), two PETIAs (sCAL turbo, GCAL), and one CLIA (QuantaFlash) and compared with a BÜHLMANN ELISA reference. ELISA assays were measured with a secondary 10-fold dilution, while PETIA and CLIA assays were performed on automated platforms with reflex dilution. Spiking experiments were conducted to evaluate recovery. RESULTS: All assays except BMA (r = 0.52) showed strong correlation with the reference (r ≥ 0.96). Median values in healthy controls varied markedly (0.29-3.22 µg/mL), with inter-assay differences exceeding one log scale. Several assays underestimated very high concentrations (> 50 µg/mL), likely due to Hook effects, and upper detection limits differed substantially. Spike recovery ranged from 79 to 268%, with the most consistent results observed for sCAL turbo. CONCLUSIONS: Commercial assays reliably quantify serum calprotectin in the normal to moderately elevated range but are not interchangeable. Variability in detection limits and susceptibility to Hook effects may lead to underestimation in severe inflammation. Assay-specific reference ranges and additional dilution steps are recommended, particularly when extreme elevations are suspected.

Molecular and Cellular PediatricsVol. 13(1)
University of Münster (DE), University Hospital of Bern (CH), University Hospital Münster (DE)
S100 Proteins and Annexins
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon