Evaluation of 24-Hour D-Dimer Stability at Room Temperature

Abstract Objective To determine the stability of plasma D-dimer after 24 hours of storage at room temperature (23–25°C). Methods In this paired ex vivo study, citrated plasma from 365 oncology patients was analyzed at baseline (HemosIL D-Dimer HS 500 on an ACL TOP 550) and again after 24 hours of room-temperature storage in capped microtubes. Agreement was assessed with the Wilcoxon signed-rank test, Passing–Bablok regression, percentage-difference Bland–Altman analysis, and diagnostic concordance at 500 ng/mL fibrinogen equivalent units (FEU) (0.50 mg/L FEU). Results Median D-dimer was 292 ng/mL FEU at baseline and 294 ng/mL FEU at 24 hours ( P = .70). Passing–Bablok regression showed no constant or proportional bias. The mean relative bias was − 1.6% (95% limits of agreement, − 45.4% to 42.3%), with the largest differences at low concentrations. Diagnostic concordance was 95.9% (350/365; κ = 0.90), but 12 of 108 baseline-positive samples (11.1%) became negative after storage. Conclusions Plasma D-dimer showed high overall stability after 24 hours at room temperature, suggesting that laboratories can process delayed samples. Borderline-positive results should be interpreted cautiously because of a small risk of false-negative classification.

Authors

Institutions

Publication Details

Journal
Research Square
Published
2026-11-10
DOI
https://doi.org/10.21203/rs.3.rs-11638370/v1
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Evaluation of 24-Hour D-Dimer Stability at Room Temperature

Usi Sukorini, Gustav Valentino
Research Square
Clinical Laboratory Practices and Quality Control
preprint

Evaluation of 24-Hour D-Dimer Stability at Room Temperature

Usi Sukorini, Gustav Valentino
preprint en

Abstract

Abstract Objective To determine the stability of plasma D-dimer after 24 hours of storage at room temperature (23–25°C). Methods In this paired ex vivo study, citrated plasma from 365 oncology patients was analyzed at baseline (HemosIL D-Dimer HS 500 on an ACL TOP 550) and again after 24 hours of room-temperature storage in capped microtubes. Agreement was assessed with the Wilcoxon signed-rank test, Passing–Bablok regression, percentage-difference Bland–Altman analysis, and diagnostic concordance at 500 ng/mL fibrinogen equivalent units (FEU) (0.50 mg/L FEU). Results Median D-dimer was 292 ng/mL FEU at baseline and 294 ng/mL FEU at 24 hours ( P = .70). Passing–Bablok regression showed no constant or proportional bias. The mean relative bias was − 1.6% (95% limits of agreement, − 45.4% to 42.3%), with the largest differences at low concentrations. Diagnostic concordance was 95.9% (350/365; κ = 0.90), but 12 of 108 baseline-positive samples (11.1%) became negative after storage. Conclusions Plasma D-dimer showed high overall stability after 24 hours at room temperature, suggesting that laboratories can process delayed samples. Borderline-positive results should be interpreted cautiously because of a small risk of false-negative classification.

Research Square
Universitas Gadjah Mada (ID), President University (ID)
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon