Interface Pressure Optimization for a Lower Limb Assistive Exoskeleton—Mechanical Modeling and Monte Carlo Simulation of Segmented Compliant Supports

The effect of the straps and support structures of a lower limb gait assist exoskeleton on pressure at the skin and soft tissue interface is examined. Because muscle atrophy may alter how loads pass between the limb and the device, a quasi-static torque and contact model for a lateral knee support model was developed. The nominal model used an assistive torque of 8 N·m and a support spacing of 0.24 m. A narrow circumferential cuff, a wide circumferential cuff, and segmented compliant pads were compared using the Markov Chain Monte Carlo method. Assistive torque, support spacing, effective contact fraction, strap preload, and a local concentration factor were varied. Under the stated model assumptions, the segmented pad concept had a median mean pressure of 8.9 kPa, with a 95% simulation interval of 6.8 to 11.5 kPa. Its peak pressure proxy had a median of 11.7 kPa, with an interval of 8.5 to 16.1 kPa. The corresponding medians for the narrow cuff were 20.6 and 36.9 kPa. Based on the comparison, an open, segmented load-bearing pad is proposed and bench testing on a limb surrogate is recommended. In the future, this study has implications on wearable pressure limits for older adults.

Authors

Publication Details

Journal
Applied and Computational Engineering
Published
2026-11-10
DOI
https://doi.org/10.54254/2755-2721/2026.38751
Primary Topic
Prosthetics and Rehabilitation Robotics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Interface Pressure Optimization for a Lower Limb Assistive Exoskeleton—Mechanical Modeling and Monte Carlo Simulation of Segmented Compliant Supports

Dongze Li
Applied and Computational Engineering
Prosthetics and Rehabilitation Robotics
article

Interface Pressure Optimization for a Lower Limb Assistive Exoskeleton—Mechanical Modeling and Monte Carlo Simulation of Segmented Compliant Supports

Dongze Li
article en

Abstract

The effect of the straps and support structures of a lower limb gait assist exoskeleton on pressure at the skin and soft tissue interface is examined. Because muscle atrophy may alter how loads pass between the limb and the device, a quasi-static torque and contact model for a lateral knee support model was developed. The nominal model used an assistive torque of 8 N·m and a support spacing of 0.24 m. A narrow circumferential cuff, a wide circumferential cuff, and segmented compliant pads were compared using the Markov Chain Monte Carlo method. Assistive torque, support spacing, effective contact fraction, strap preload, and a local concentration factor were varied. Under the stated model assumptions, the segmented pad concept had a median mean pressure of 8.9 kPa, with a 95% simulation interval of 6.8 to 11.5 kPa. Its peak pressure proxy had a median of 11.7 kPa, with an interval of 8.5 to 16.1 kPa. The corresponding medians for the narrow cuff were 20.6 and 36.9 kPa. Based on the comparison, an open, segmented load-bearing pad is proposed and bench testing on a limb surrogate is recommended. In the future, this study has implications on wearable pressure limits for older adults.

Applied and Computational EngineeringVol. 278(1)
Prosthetics and Rehabilitation Robotics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon