A Low-Cost Pulse Oximeter with an Honest Accuracy Budget: Design, Error Propagation, and the Accuracy Floor for Triage

Design study (simulated, not measured) of a transmissive finger-clip pulse oximeter built from discrete low-cost parts (bill of materials US $17.50). The ratio-of-ratios response is derived from published haemoglobin extinction coefficients at 660 nm and 940 nm, and detector noise, residual baseline disturbance, motion artefact, calibration scatter, and a melanin-dependent bias term are propagated into a total root-mean-square accuracy (A_rms) over 70-100% saturation as a function of perfusion index, averaging time, motion, and skin pigmentation. Computed A_rms is 1.52 percentage points at perfusion index 3% at rest in light/medium pigmentation, 3.37 points in dark pigmentation with a broad emitter and single calibration, and 2.41 points at the dark low-perfusion corner after narrowband emitters, melanin-stratified calibration, and a 16 s window; raising LED drive current does not improve accuracy. Below perfusion index 1%, or where A_rms exceeds 3.0 points in the conditions of use, the device must blank its readings and must not be trusted for triage.

Authors

Publication Details

Journal
Zenodo (CERN European Organization for Nuclear Research)
Published
2026-11-10
DOI
https://doi.org/10.5281/zenodo.24108849
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

A Low-Cost Pulse Oximeter with an Honest Accuracy Budget: Design, Error Propagation, and the Accuracy Floor for Triage

William Alexander Aldred
Zenodo (CERN European Organization for Nuclear Research)
Non-Invasive Vital Sign Monitoring
article

A Low-Cost Pulse Oximeter with an Honest Accuracy Budget: Design, Error Propagation, and the Accuracy Floor for Triage

William Alexander Aldred
article en

Abstract

Design study (simulated, not measured) of a transmissive finger-clip pulse oximeter built from discrete low-cost parts (bill of materials US $17.50). The ratio-of-ratios response is derived from published haemoglobin extinction coefficients at 660 nm and 940 nm, and detector noise, residual baseline disturbance, motion artefact, calibration scatter, and a melanin-dependent bias term are propagated into a total root-mean-square accuracy (A_rms) over 70-100% saturation as a function of perfusion index, averaging time, motion, and skin pigmentation. Computed A_rms is 1.52 percentage points at perfusion index 3% at rest in light/medium pigmentation, 3.37 points in dark pigmentation with a broad emitter and single calibration, and 2.41 points at the dark low-perfusion corner after narrowband emitters, melanin-stratified calibration, and a 16 s window; raising LED drive current does not improve accuracy. Below perfusion index 1%, or where A_rms exceeds 3.0 points in the conditions of use, the device must blank its readings and must not be trusted for triage.

Zenodo (CERN European Organization for Nuclear Research)
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon