Validation of an LC-Q-TOF “HiRes Mass-fix” assay for improved sensitivity and specificity of M-protein detection

Abstract Objectives To validate a blood-based liquid chromatography quadrupole time-of-flight mass spectrometry (MS) method “HiRes Mass-Fix” as a fit-for-purpose, patient-specific method for detecting low level M-proteins in patients who have reached negativity by MALDI-TOF MS “Mass-Fix”. Methods Analytical validation included precision, accuracy, analytical measurement interval, limit of detection, specificity, ion suppression, carryover, and extracted-sample stability. Clinical performance was assessed by comparing 174 patients with bone marrow (BM) measurable residual disease (MRD) by next-generation flow cytometry (NGF) to both Mass-Fix and HiRes Mass-Fix. Results HiRes Mass-Fix demonstrated acceptable precision, accuracy, linearity, specificity, and stability for its intended use, with an average limit of detection of 3.8 mg/L. HiRes Mass-Fix detected M-proteins in 41 of 58 Mass-Fix negative samples. This raised blood MRD positivity from 62 to 89 % against a BM NGF positivity rate of 51 %. Blood-BM agreement fell from 71 % (kappa 0.43) by Mass-Fix to 58 % (kappa 0.15) by HiRes Mass-Fix, driven by blood-positive/BM-negative discordance. IgG was overrepresented among discordant cases (68 vs. 45 %, p=0.003). Conclusions HiRes Mass-Fix was successfully validated and demonstrated to increase M-protein detection in samples negative by Mass-Fix. IgG recycling may explain the blood-positive/BM-negative discordance, and serial measurements will be needed to distinguish residual M-protein from ongoing production. Blood-based M-protein measurement is a less invasive, lower-cost alternative to bone marrow biopsy for ongoing patient monitoring.

Authors

Institutions

Publication Details

Journal
Clinical Chemistry and Laboratory Medicine (CCLM)
Published
2026-11-10
DOI
https://doi.org/10.1515/cclm-2026-1525
Primary Topic
Multiple Myeloma Research and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Validation of an LC-Q-TOF “HiRes Mass-fix” assay for improved sensitivity and specificity of M-protein detection

Ryan W. Pearce, Dragan Jevremović, Maria Alice V. Willrich, Mindy Kohlhagen et al.
Clinical Chemistry and Laboratory Medicine (CCLM)
Multiple Myeloma Research and Treatments
article

Validation of an LC-Q-TOF “HiRes Mass-fix” assay for improved sensitivity and specificity of M-protein detection

Ryan W. Pearce, Dragan Jevremović, Maria Alice V. Willrich, Mindy Kohlhagen, David L. Murray, Jonathan Coker, Mark A. Martinez
article en

Abstract

Abstract Objectives To validate a blood-based liquid chromatography quadrupole time-of-flight mass spectrometry (MS) method “HiRes Mass-Fix” as a fit-for-purpose, patient-specific method for detecting low level M-proteins in patients who have reached negativity by MALDI-TOF MS “Mass-Fix”. Methods Analytical validation included precision, accuracy, analytical measurement interval, limit of detection, specificity, ion suppression, carryover, and extracted-sample stability. Clinical performance was assessed by comparing 174 patients with bone marrow (BM) measurable residual disease (MRD) by next-generation flow cytometry (NGF) to both Mass-Fix and HiRes Mass-Fix. Results HiRes Mass-Fix demonstrated acceptable precision, accuracy, linearity, specificity, and stability for its intended use, with an average limit of detection of 3.8 mg/L. HiRes Mass-Fix detected M-proteins in 41 of 58 Mass-Fix negative samples. This raised blood MRD positivity from 62 to 89 % against a BM NGF positivity rate of 51 %. Blood-BM agreement fell from 71 % (kappa 0.43) by Mass-Fix to 58 % (kappa 0.15) by HiRes Mass-Fix, driven by blood-positive/BM-negative discordance. IgG was overrepresented among discordant cases (68 vs. 45 %, p=0.003). Conclusions HiRes Mass-Fix was successfully validated and demonstrated to increase M-protein detection in samples negative by Mass-Fix. IgG recycling may explain the blood-positive/BM-negative discordance, and serial measurements will be needed to distinguish residual M-protein from ongoing production. Blood-based M-protein measurement is a less invasive, lower-cost alternative to bone marrow biopsy for ongoing patient monitoring.

Clinical Chemistry and Laboratory Medicine (CCLM)
Mayo Clinic (US)
Multiple Myeloma Research and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon