SERS–LFIA technology based on Au@Ag nanospheres for simultaneous detection of carbapenemases KPC and NDM
. The coefficients of variation for repeatability testing at low, medium, and high antigen concentrations were all less than 15%. The cut-off values for KPC and NDM were determined to be 846 and 784, respectively. Using PCR results as the gold standard, the diagnostic sensitivity and specificity of this method were both 100%, demonstrating a 100% agreement rate with commercial colloidal gold detection kits. The SERS-LFIA detection system may provide technical support for rapid identification of carbapenemase types.
Authors
- Chunyan Liu (ORCID: https://orcid.org/0009-0002-7145-4870)
- Qian Shi (ORCID: https://orcid.org/0000-0002-4640-8832)
- Mengjie Liang (ORCID: https://orcid.org/0000-0003-2932-9830)
- Hefei Zha
- Xueli Zhang
- Xin Zhang (ORCID: https://orcid.org/0009-0004-8552-8089)
Institutions
- Shihezi University (CN)
- Xinjiang Production and Construction Corps (CN)
Publication Details
- Journal
- RSC Advances
- Published
- 2026-11-18
- DOI
- https://doi.org/10.1039/d6ra08505a
- Primary Topic
- Biosensors and Analytical Detection
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Xinjiang Production and Construction Corps