Needle Detection software improves procedural efficiency and reduces verification scans in CT-guided needle placement: a multi-operator phantom study

Abstract Purpose To assess the impact of a dedicated Needle Detection software tool on procedural efficiency and radiation-related surrogate parameters during CT-guided needle placement in a controlled, multi-operator phantom study. Materials and methods Ten radiologists performed a total of 60 CT-guided needle insertions in a torso phantom using a randomized, balanced study design (30 procedures with Needle Detection activated and 30 without). Needle Detection software is an automated image-processing algorithm that identifies, enhances, and continuously tracks the needle shaft and tip on CT images, thereby improving visual guidance during needle advancement. Primary outcomes were time to target and verification scans, defined as the number of CT verification scans performed during needle progression. Time to target was compared using the Mann–Whitney U test, while number of verification scans was analyzed using a permutation test with 10,000 iterations. Results Needle Detection significantly improved procedural performance. Mean time to target decreased from 422.5 ± 199.8 s without Needle Detection to 224.3 ± 103.3 s with Needle Detection, corresponding to an approximate 47% reduction in procedure time ( p < 0.001). The number of verification scans required during needle advancement was also significantly reduced, with a median of 4 (IQR: 3–6) without Needle Detection compared with 3 (IQR: 2–4) with Needle Detection ( p < 0.0001). Improvements in both procedural time and verification scans were observed consistently across all participating operators. Conclusion Needle Detection software significantly improved procedural efficiency during CT-guided needle placement in this randomized phantom study and was associated with fewer verification scans. Because radiation dose was not measured directly, these findings should not be interpreted as evidence of reduced patient radiation dose. Further clinical studies including direct radiation dosimetry are warranted.

Authors

Institutions

Publication Details

Journal
CVIR Oncology
Published
2026-11-09
DOI
https://doi.org/10.1007/s44343-026-00071-1
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Needle Detection software improves procedural efficiency and reduces verification scans in CT-guided needle placement: a multi-operator phantom study

Saher Saeed, Denise Schoenbeck, Jan Borggrefe, Maria Katz et al.
CVIR Oncology
Radiation Dose and Imaging
article

Needle Detection software improves procedural efficiency and reduces verification scans in CT-guided needle placement: a multi-operator phantom study

Saher Saeed, Denise Schoenbeck, Jan Borggrefe, Maria Katz, Julian Schreck, Alexander Mendorf, Julius Henning Niehoff, M. Ramona, Christoph Moenninghoff, Jan Robert Kroeger, Jonas Ottemoeller, Dyar Shekho, Thomas Roettcher, Jan Lueking, Mohammad Almatar
article en

Abstract

Abstract Purpose To assess the impact of a dedicated Needle Detection software tool on procedural efficiency and radiation-related surrogate parameters during CT-guided needle placement in a controlled, multi-operator phantom study. Materials and methods Ten radiologists performed a total of 60 CT-guided needle insertions in a torso phantom using a randomized, balanced study design (30 procedures with Needle Detection activated and 30 without). Needle Detection software is an automated image-processing algorithm that identifies, enhances, and continuously tracks the needle shaft and tip on CT images, thereby improving visual guidance during needle advancement. Primary outcomes were time to target and verification scans, defined as the number of CT verification scans performed during needle progression. Time to target was compared using the Mann–Whitney U test, while number of verification scans was analyzed using a permutation test with 10,000 iterations. Results Needle Detection significantly improved procedural performance. Mean time to target decreased from 422.5 ± 199.8 s without Needle Detection to 224.3 ± 103.3 s with Needle Detection, corresponding to an approximate 47% reduction in procedure time ( p < 0.001). The number of verification scans required during needle advancement was also significantly reduced, with a median of 4 (IQR: 3–6) without Needle Detection compared with 3 (IQR: 2–4) with Needle Detection ( p < 0.0001). Improvements in both procedural time and verification scans were observed consistently across all participating operators. Conclusion Needle Detection software significantly improved procedural efficiency during CT-guided needle placement in this randomized phantom study and was associated with fewer verification scans. Because radiation dose was not measured directly, these findings should not be interpreted as evidence of reduced patient radiation dose. Further clinical studies including direct radiation dosimetry are warranted.

CVIR OncologyVol. 2(1)
Johannes Wesling Klinikum Minden (DE), Ruhr University Bochum (DE)
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon