Surrogate-assisted multi-objective optimization of an axial-flow ventricular assist device rotor using a large CFD-derived design space

Surrogate-assisted optimization can reduce the cost of exploring ventricular-assist-device (VAD) rotor geometries, but the optimization variables, physical constraints, and uncertainty of the surrogate predictions must be treated consistently. This study analyzes a CFD-derived design space of 4809 axial-flow VAD rotors defined by six geometric variables at 5 L/min and 10,000 rpm. Seven archived responses were modeled using output-specific Random Forest, Extra Trees, Gradient Boosting, and radial-basis-function support-vector regressors. Independent test-set R 2 values ranged from 0.964 to 0.994. Hydraulic efficiency and the archived exposure-time output were retained as diagnostic responses but excluded from optimization because consistency audits did not support their use as independent physical objectives. A constraint-aware multi-objective particle swarm optimization therefore minimized torque, hemolysis index, and platelet activation state and maximized the volume fraction below 50 Pa, while requiring pressure rise to meet the baseline target. Five runs with independent seeds were combined through their best-hypervolume archives, and five diverse Pareto candidates were re-evaluated by direct CFD. The final CFD-confirmed compromise used four blades, a 52.18° inlet angle, a 25.00° outlet angle, a 0.105 mm clearance, a 0.600 mm blade thickness, and a 38.30 mm rotor length. Relative to the baseline, direct CFD maintained pressure rise (+0.15%), reduced torque by 6.77%, increased the low-shear volume fraction by 4.14%, and reduced the comparative hemolysis and platelet-activation indicators by 36.07% and 26.38%, respectively. The results establish a reproducible fixed-operating-point design-screening workflow, while absolute blood-damage interpretation and operation beyond the prescribed point require further validation.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2026-11-08
DOI
https://doi.org/10.1177/09544119261504129
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Surrogate-assisted multi-objective optimization of an axial-flow ventricular assist device rotor using a large CFD-derived design space

Wajih Maazouzi, Reda Lakraimi, Mourad Taha‐Janan, Mohamed Bounouib et al.
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Mechanical Circulatory Support Devices
article

Surrogate-assisted multi-objective optimization of an axial-flow ventricular assist device rotor using a large CFD-derived design space

Wajih Maazouzi, Reda Lakraimi, Mourad Taha‐Janan, Mohamed Bounouib, Kamal Bouzid, Hamza Isksioui
article en

Abstract

Surrogate-assisted optimization can reduce the cost of exploring ventricular-assist-device (VAD) rotor geometries, but the optimization variables, physical constraints, and uncertainty of the surrogate predictions must be treated consistently. This study analyzes a CFD-derived design space of 4809 axial-flow VAD rotors defined by six geometric variables at 5 L/min and 10,000 rpm. Seven archived responses were modeled using output-specific Random Forest, Extra Trees, Gradient Boosting, and radial-basis-function support-vector regressors. Independent test-set R 2 values ranged from 0.964 to 0.994. Hydraulic efficiency and the archived exposure-time output were retained as diagnostic responses but excluded from optimization because consistency audits did not support their use as independent physical objectives. A constraint-aware multi-objective particle swarm optimization therefore minimized torque, hemolysis index, and platelet activation state and maximized the volume fraction below 50 Pa, while requiring pressure rise to meet the baseline target. Five runs with independent seeds were combined through their best-hypervolume archives, and five diverse Pareto candidates were re-evaluated by direct CFD. The final CFD-confirmed compromise used four blades, a 52.18° inlet angle, a 25.00° outlet angle, a 0.105 mm clearance, a 0.600 mm blade thickness, and a 38.30 mm rotor length. Relative to the baseline, direct CFD maintained pressure rise (+0.15%), reduced torque by 6.77%, increased the low-shear volume fraction by 4.14%, and reduced the comparative hemolysis and platelet-activation indicators by 36.07% and 26.38%, respectively. The results establish a reproducible fixed-operating-point design-screening workflow, while absolute blood-damage interpretation and operation beyond the prescribed point require further validation.

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Mohammed V University (MA), Arts et Métiers (FR), Abdelmalek Essaâdi University (MA), Université Moulay Ismail de Meknes (MA)
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon