Effects of Ambulatory Blood Pressure Monitoring During Polysomnography on Sleep Indices and EEG Power

Background/Objectives: Ambulatory blood pressure monitoring (ABPM) may disturb sleep during polysomnography (PSG), and the first-night effect (FNE) can complicate interpretation. We evaluated changes associated with ABPM and FNE across whole nights and individual measurements. Methods: This secondary analysis included 72 nominally healthy adults in a balanced two-night crossover, with ABPM on one randomly assigned night. Linear mixed-effects models estimated associations with PSG-derived sleep indices, a composite sleep quality index (SQI), and relative electroencephalographic (EEG) power. Analyses of individual measurements summarized EEG changes, sleep-stage transitions, and automated arousal detections. Benjamini–Hochberg false discovery rate (BH-FDR) adjustment was applied within defined analysis families. Results: FNE and ABPM were associated with lower SQI (β = −0.049 and −0.051, respectively; both BH-adjusted p = 0.028). FNE was associated with altered sleep continuity and architecture, whereas ABPM was associated with higher light-sleep and lower N3-sleep percentages and higher whole-night relative theta power. In the Night 2 paired analysis, relative delta power was lower and relative alpha, beta, and gamma power were higher during ABPM than before ABPM (all BH-adjusted p < 0.01); the theta difference did not reach significance (BH-adjusted p = 0.0632). Of 655 eligible ABPM measurements beginning during sleep, 395 (60.3%) contained at least one arousal detection within the 2 min measurement window. Conclusions: These detections indicate temporal co-occurrence and do not establish that cuff inflation caused arousal. Concurrent ABPM and study-night order should be considered when interpreting PSG. An adaptation PSG night may be considered when feasible, as may cuffless blood pressure monitoring once its accuracy has been established for the intended setting.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2026-11-09
DOI
https://doi.org/10.3390/jcm15218759
Primary Topic
Sleep and related disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Effects of Ambulatory Blood Pressure Monitoring During Polysomnography on Sleep Indices and EEG Power

Jinseok Park, Jeyeon Lee, Myung-Kyu Yi, Byung Sik Kim et al.
Journal of Clinical Medicine
Sleep and related disorders
article

Effects of Ambulatory Blood Pressure Monitoring During Polysomnography on Sleep Indices and EEG Power

Jinseok Park, Jeyeon Lee, Myung-Kyu Yi, Byung Sik Kim, Jeong‐Hun Shin, Jongshill Lee, J.K. Kim, Hyun‐Jin Kim, In Young Kim, Deok-Gyeong Kang, Junhee Han
article en

Abstract

Background/Objectives: Ambulatory blood pressure monitoring (ABPM) may disturb sleep during polysomnography (PSG), and the first-night effect (FNE) can complicate interpretation. We evaluated changes associated with ABPM and FNE across whole nights and individual measurements. Methods: This secondary analysis included 72 nominally healthy adults in a balanced two-night crossover, with ABPM on one randomly assigned night. Linear mixed-effects models estimated associations with PSG-derived sleep indices, a composite sleep quality index (SQI), and relative electroencephalographic (EEG) power. Analyses of individual measurements summarized EEG changes, sleep-stage transitions, and automated arousal detections. Benjamini–Hochberg false discovery rate (BH-FDR) adjustment was applied within defined analysis families. Results: FNE and ABPM were associated with lower SQI (β = −0.049 and −0.051, respectively; both BH-adjusted p = 0.028). FNE was associated with altered sleep continuity and architecture, whereas ABPM was associated with higher light-sleep and lower N3-sleep percentages and higher whole-night relative theta power. In the Night 2 paired analysis, relative delta power was lower and relative alpha, beta, and gamma power were higher during ABPM than before ABPM (all BH-adjusted p < 0.01); the theta difference did not reach significance (BH-adjusted p = 0.0632). Of 655 eligible ABPM measurements beginning during sleep, 395 (60.3%) contained at least one arousal detection within the 2 min measurement window. Conclusions: These detections indicate temporal co-occurrence and do not establish that cuff inflation caused arousal. Concurrent ABPM and study-night order should be considered when interpreting PSG. An adaptation PSG night may be considered when feasible, as may cuffless blood pressure monitoring once its accuracy has been established for the intended setting.

Journal of Clinical MedicineVol. 15(21)
Hanyang University Guri Hospital (KR), Hanyang University Seoul Hospital (KR), Hanyang University (KR)
Sleep and related disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon