Co-localized OCT and diffuse reflectance spectroscopy in a fiber-optic probe for dosimetry in retinal endolaser photocoagulation

Retinal endolaser photocoagulation (REPC) is a widely performed vitreoretinal procedure, yet quantitative intraoperative treatment outcome assessment remains challenging using conventional operating microscopes or microscope-integrated optical coherence tomography (miOCT) systems. In this work, we present a dual-fiber 21G endolaser probe with instrument-integrated optical coherence tomography (iiOCT) and diffuse reflectance spectroscopy (DRS) via a monolithic 3D nano-printed sensor head for co-localized laser delivery and multimodal sensing. Phase-sensitive iiOCT M-scans (consecutive A-scans acquired over time) monitor structural tissue alterations during laser irradiation, while broadband illumination and collection via a dual-clad fiber and multi-mode fiber capture pre- and post-operative diffuse reflectance spectra. Using ex vivo porcine eyes, sensor-derived metrics were evaluated against microscopic lesion visibility grades and miOCT-measured lesion diameters using mixed-effects models under leave-one-eye-out cross-validation. Both single-sensor channels improved predictive performance over laser parameters alone. For ordinal lesion visibility classification, the DRS-derived spectral tilt increased overall accuracy from 0.784 to 0.867 and the correct classification rate for barely visible lesions from 38.1% to 66.7%. For lesion diameter prediction, the iiOCT-derived metric achieved the best single-sensor performance, increasing mean R² by 5.7 percentage points to 0.776. Combining structural and spectral metrics yielded the highest performance for both tasks, with a visibility accuracy of 0.869 and a diameter-prediction R² of 0.790. These results demonstrate the feasibility of co-localized multimodal sensing in a single fiber-optic probe and provide a foundation for future real-time, sensor-guided endolaser surgery.

Authors

Publication Details

Published
2026-11-09
DOI
https://doi.org/10.1364/opticaopen.34936863.v1
Primary Topic
Optical Coherence Tomography Applications
Type
preprint
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
preprint

Co-localized OCT and diffuse reflectance spectroscopy in a fiber-optic probe for dosimetry in retinal endolaser photocoagulation

Philipp Matten, Ç‪ağlar Ataman, Florian Lux, Nicklas Enenkel et al.
Optical Coherence Tomography Applications
preprint

Co-localized OCT and diffuse reflectance spectroscopy in a fiber-optic probe for dosimetry in retinal endolaser photocoagulation

Philipp Matten, Ç‪ağlar Ataman, Florian Lux, Nicklas Enenkel, Ian Howard, Pragyamita Basu, Dongyue Wu
preprint en

Abstract

Retinal endolaser photocoagulation (REPC) is a widely performed vitreoretinal procedure, yet quantitative intraoperative treatment outcome assessment remains challenging using conventional operating microscopes or microscope-integrated optical coherence tomography (miOCT) systems. In this work, we present a dual-fiber 21G endolaser probe with instrument-integrated optical coherence tomography (iiOCT) and diffuse reflectance spectroscopy (DRS) via a monolithic 3D nano-printed sensor head for co-localized laser delivery and multimodal sensing. Phase-sensitive iiOCT M-scans (consecutive A-scans acquired over time) monitor structural tissue alterations during laser irradiation, while broadband illumination and collection via a dual-clad fiber and multi-mode fiber capture pre- and post-operative diffuse reflectance spectra. Using ex vivo porcine eyes, sensor-derived metrics were evaluated against microscopic lesion visibility grades and miOCT-measured lesion diameters using mixed-effects models under leave-one-eye-out cross-validation. Both single-sensor channels improved predictive performance over laser parameters alone. For ordinal lesion visibility classification, the DRS-derived spectral tilt increased overall accuracy from 0.784 to 0.867 and the correct classification rate for barely visible lesions from 38.1% to 66.7%. For lesion diameter prediction, the iiOCT-derived metric achieved the best single-sensor performance, increasing mean R² by 5.7 percentage points to 0.776. Combining structural and spectral metrics yielded the highest performance for both tasks, with a visibility accuracy of 0.869 and a diameter-prediction R² of 0.790. These results demonstrate the feasibility of co-localized multimodal sensing in a single fiber-optic probe and provide a foundation for future real-time, sensor-guided endolaser surgery.

Optical Coherence Tomography Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon
Co-localized OCT and diffuse reflectance spectroscopy in a fiber-optic probe for dosimetry in retinal endolaser photocoagulation — Philipp Matten, Ç‪ağlar Ataman, et al. · (2026) | TGRS Research Map | TGRS