Warm, ventilate, stabilize: Clinical feasibility and non-inferiority study of MINAR™ infant warmer resuscitator

Background: Effective neonatal resuscitation requires integrated thermal support, respiratory assistance, and continuous physiological monitoring. MINAR™, a locally manufactured infant warmer resuscitator developed in Indonesia, was designed to support neonatal stabilization. This study evaluated the clinical feasibility, safety, and non-inferiority of MINAR™ compared with the Dräger Resuscitaire ® . Methods: This single-center, randomized, non-inferiority feasibility trial allocated live-born neonates requiring postnatal resuscitation or stabilization in a 1:1 ratio to MINAR™ or Dräger Resuscitaire ® . Primary outcomes included respiratory rate (RR), heart rate (HR), peripheral oxygen saturation (SpO 2 ), body temperature, Apgar score, and Downes score measured at baseline and 5, 10, 15, and 60 min. Minimum clinically important differences were predefined. Repeated-measures ANCOVA was performed to assess the potential confounding effect of baseline Downes score. Results: Forty-eight neonates were enrolled, and 38 were included in the final analysis (MINAR™, n = 18; Dräger Resuscitaire ® , n = 20). Baseline characteristics were comparable except for a higher baseline Downes score in the Dräger group ( p = 0.004). MINAR™ demonstrated non-inferior performance across all predefined outcomes with no differences exceeding the non-inferiority margins. After adjustment for baseline respiratory distress, MINAR™ remained associated with faster improvement in oxygen saturation and greater reduction in Downes score, whereas RR, HR, body temperature, and Apgar score showed comparable longitudinal trajectories between groups. No device-related adverse events or malfunctions were observed. Conclusions: MINAR™ demonstrated feasibility, safety, and non-inferior clinical performance compared with the Dräger Resuscitaire ® for neonatal postnatal stabilization. The observed advantages in oxygenation and respiratory distress remained independent of baseline respiratory severity, supporting further evaluation in larger multicenter randomized trials. Clinical Trial Registry: ISRCTN88734666

Authors

Institutions

Publication Details

Journal
Journal of the Intensive Care Society
Published
2026-11-08
DOI
https://doi.org/10.1177/17511437261497940
Primary Topic
Neonatal Respiratory Health Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Warm, ventilate, stabilize: Clinical feasibility and non-inferiority study of MINAR™ infant warmer resuscitator

Lia Nazliah Abdullah, Reza Widianto Sudjud, Jenifer Kiem Aviani
Journal of the Intensive Care Society
Neonatal Respiratory Health Research
article

Warm, ventilate, stabilize: Clinical feasibility and non-inferiority study of MINAR™ infant warmer resuscitator

Lia Nazliah Abdullah, Reza Widianto Sudjud, Jenifer Kiem Aviani
article en

Abstract

Background: Effective neonatal resuscitation requires integrated thermal support, respiratory assistance, and continuous physiological monitoring. MINAR™, a locally manufactured infant warmer resuscitator developed in Indonesia, was designed to support neonatal stabilization. This study evaluated the clinical feasibility, safety, and non-inferiority of MINAR™ compared with the Dräger Resuscitaire ® . Methods: This single-center, randomized, non-inferiority feasibility trial allocated live-born neonates requiring postnatal resuscitation or stabilization in a 1:1 ratio to MINAR™ or Dräger Resuscitaire ® . Primary outcomes included respiratory rate (RR), heart rate (HR), peripheral oxygen saturation (SpO 2 ), body temperature, Apgar score, and Downes score measured at baseline and 5, 10, 15, and 60 min. Minimum clinically important differences were predefined. Repeated-measures ANCOVA was performed to assess the potential confounding effect of baseline Downes score. Results: Forty-eight neonates were enrolled, and 38 were included in the final analysis (MINAR™, n = 18; Dräger Resuscitaire ® , n = 20). Baseline characteristics were comparable except for a higher baseline Downes score in the Dräger group ( p = 0.004). MINAR™ demonstrated non-inferior performance across all predefined outcomes with no differences exceeding the non-inferiority margins. After adjustment for baseline respiratory distress, MINAR™ remained associated with faster improvement in oxygen saturation and greater reduction in Downes score, whereas RR, HR, body temperature, and Apgar score showed comparable longitudinal trajectories between groups. No device-related adverse events or malfunctions were observed. Conclusions: MINAR™ demonstrated feasibility, safety, and non-inferior clinical performance compared with the Dräger Resuscitaire ® for neonatal postnatal stabilization. The observed advantages in oxygenation and respiratory distress remained independent of baseline respiratory severity, supporting further evaluation in larger multicenter randomized trials. Clinical Trial Registry: ISRCTN88734666

Journal of the Intensive Care Society
Dr. Hasan Sadikin General Hospital (ID), Padjadjaran University (ID)
Neonatal Respiratory Health Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon