Initial Force Expression of Different Clear Aligner Systems in a Simulated Maxillary Canine Distalization Model with Lateral Incisor Agenesis: An In Vitro Comparative Study

Background: This in vitro study compared initial force release by different clear aligner systems during simulated maxillary canine distalization with lateral incisor agenesis and evaluated attachment effects. Methods: Five aligner systems were tested: SmartTrackTM (Invisalign®, Align Technology Inc., Santa Clara, CA, USA), TruGENTM (Spark®, Ormco Corporation, Brea, CA, USA), Essix® (SureSmile®, Dentsply Sirona Inc., Charlotte, NC, USA), MasterControl S (Angel Aligner®, Angelalign Technology Inc., Shanghai, China), and a directly 3D-printed aligner fabricated with TC-85 DAC resin (Graphy®, Graphy Inc., Seoul, Republic of Korea). Ten identical aligners per configuration were tested for 1 h at 37 °C in physiological solution. The ElectroForce® system measured the translational force at the canine’s root-level support. Data were analyzed using descriptive statistics and two-way ANOVA. Results: TruGENTM with attachments generated the highest mean force (1.25 N), followed by SmartTrackTM (0.61 N), Essix® (0.42 N), MasterControl S (0.39 N), and TC-85 DAC resin (0.01 N). Without attachments, forces were minimal for all systems. Force decay was observed in groups expressing measurable forces. Conclusions: Attachments enabled measurable translational force components in thermoformed systems. Force expression differed among systems, while TC-85 DAC resin produced negligible forces with or without attachments under experimental conditions. Further clinical validation is warranted.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-11-08
DOI
https://doi.org/10.3390/bioengineering13111326
Primary Topic
Orthodontics and Dentofacial Orthopedics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Initial Force Expression of Different Clear Aligner Systems in a Simulated Maxillary Canine Distalization Model with Lateral Incisor Agenesis: An In Vitro Comparative Study

Andrea Tancredi Lugas, Nicola Scotti, Mara Terzini, Cristina Bignardi et al.
Bioengineering
Orthodontics and Dentofacial Orthopedics
article

Initial Force Expression of Different Clear Aligner Systems in a Simulated Maxillary Canine Distalization Model with Lateral Incisor Agenesis: An In Vitro Comparative Study

Andrea Tancredi Lugas, Nicola Scotti, Mara Terzini, Cristina Bignardi, Claudia Dettori, Enrico Spinas, Teresa Cobo, Alessio Verdecchia, Samuele Avolese
article en

Abstract

Background: This in vitro study compared initial force release by different clear aligner systems during simulated maxillary canine distalization with lateral incisor agenesis and evaluated attachment effects. Methods: Five aligner systems were tested: SmartTrackTM (Invisalign®, Align Technology Inc., Santa Clara, CA, USA), TruGENTM (Spark®, Ormco Corporation, Brea, CA, USA), Essix® (SureSmile®, Dentsply Sirona Inc., Charlotte, NC, USA), MasterControl S (Angel Aligner®, Angelalign Technology Inc., Shanghai, China), and a directly 3D-printed aligner fabricated with TC-85 DAC resin (Graphy®, Graphy Inc., Seoul, Republic of Korea). Ten identical aligners per configuration were tested for 1 h at 37 °C in physiological solution. The ElectroForce® system measured the translational force at the canine’s root-level support. Data were analyzed using descriptive statistics and two-way ANOVA. Results: TruGENTM with attachments generated the highest mean force (1.25 N), followed by SmartTrackTM (0.61 N), Essix® (0.42 N), MasterControl S (0.39 N), and TC-85 DAC resin (0.01 N). Without attachments, forces were minimal for all systems. Force decay was observed in groups expressing measurable forces. Conclusions: Attachments enabled measurable translational force components in thermoformed systems. Force expression differed among systems, while TC-85 DAC resin produced negligible forces with or without attachments under experimental conditions. Further clinical validation is warranted.

BioengineeringVol. 13(11)
Universidad de Oviedo (ES), University of Cagliari (IT), Politecnico di Torino (IT), University of Turin (IT)
Orthodontics and Dentofacial Orthopedics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon