Segmented G(E) Function Compensation Method for Gamma Dose Rate Meters Considering Asymmetric Error Distribution Across Energy Regions

Gamma dose rate meters are used in environmental radiation monitoring and radiation safety assessment. Conventional global G(E) functions can concentrate dose conversion errors at low energies. A segmented G(E) compensation method based on the multi-segment least-squares method (MSLSM) was developed for a compact CsI(Tl)–silicon photomultiplier (SiPM) meter, with separate functions for 20–200 keV and 200 keV < E ≤ 2000 keV. Full-spectrum fitting mean absolute percentage error (MAPE) decreased from 10.36% to 5.83%, a relative reduction of 43.73%, while low-energy MAPE decreased from 22.32% to 11.42%. For simulated nuclide spectra, the 241Am relative deviation decreased from 9.17% to 1.83%; results for 137Cs, 60Co, and most mixtures were comparable to or slightly better than those of the conventional global least-squares method (LSM). After correction, the maximum absolute relative deviations between the prototype meter and the commercial SAM945 instrument were approximately 7.5% for 137Cs and 14.91% for 60Co. Segmented G(E) modeling thus improves the regional error distribution and dose rate fitting accuracy over a wide energy range for the detector studied.

Authors

Institutions

Publication Details

Journal
Symmetry
Published
2026-11-09
DOI
https://doi.org/10.3390/sym18111874
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

Segmented G(E) Function Compensation Method for Gamma Dose Rate Meters Considering Asymmetric Error Distribution Across Energy Regions

Lei Zhong, Wenhao Ding, Rui Hu, Sheng Lv et al.
Symmetry
Radiation Detection and Scintillator Technologies
article

Segmented G(E) Function Compensation Method for Gamma Dose Rate Meters Considering Asymmetric Error Distribution Across Energy Regions

Lei Zhong, Wenhao Ding, Rui Hu, Sheng Lv, Yongjian Chen, Qing Li, Fei Li
article en

Abstract

Gamma dose rate meters are used in environmental radiation monitoring and radiation safety assessment. Conventional global G(E) functions can concentrate dose conversion errors at low energies. A segmented G(E) compensation method based on the multi-segment least-squares method (MSLSM) was developed for a compact CsI(Tl)–silicon photomultiplier (SiPM) meter, with separate functions for 20–200 keV and 200 keV < E ≤ 2000 keV. Full-spectrum fitting mean absolute percentage error (MAPE) decreased from 10.36% to 5.83%, a relative reduction of 43.73%, while low-energy MAPE decreased from 22.32% to 11.42%. For simulated nuclide spectra, the 241Am relative deviation decreased from 9.17% to 1.83%; results for 137Cs, 60Co, and most mixtures were comparable to or slightly better than those of the conventional global least-squares method (LSM). After correction, the maximum absolute relative deviations between the prototype meter and the commercial SAM945 instrument were approximately 7.5% for 137Cs and 14.91% for 60Co. Segmented G(E) modeling thus improves the regional error distribution and dose rate fitting accuracy over a wide energy range for the detector studied.

SymmetryVol. 18(11)
Chengdu University of Technology (CN), Shanghai Nuclear Engineering Research and Design Institute (China) (CN)
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon